|
Interpreting SCUBA noise performance
Interpreting SCUBA noise observations |
|
This page is designed to assist in the interpretation of SCUBA noise measurements and the diagnosis
of problems associated with excess noise levels.
1. Introduction: Why take SCUBA noise measurements?
In most cases the first thing to check, once the TCS and SCUBA software is up and running, is that SCUBA
is performing well. This is done in the first instance by the NOISE observation. The observation
definition file is called noise_ref.obs (usually it is contained in the startup.m macro) and measures the
"system" noise for each pixel looking at a cold load (actually a reflecting mirror in front of the
cryostat window). Removing the mirror allows the "sky" noise level to be measured (noise_sky.obs) which is
useful indicator of the stability of observing conditions. The rest of this report is mainly concerned
with the interpretation of system noise data, and how to diagnose problems when excess noise levels are
encountered.
2. Interpreting noise observations
A noise observation will display the rms noise levels for each pixel after 64 seconds of integration time.
The on-line display takes the form of a plot of noise voltage on the y-axis against pixel number on the
x-axis. Since this display will autoscale dependent on the most noisy pixel, it is often advantageous to
run the SCUNOISE utility (part of OBSDESK) after the observation has completed to look more closely at the
mean noise level and identify where the noisy pixels are on the arrays.
The case shown below is an example of a good noise performance (excellent in fact!). On the left is a plot
of noise voltage against pixel number. The first 91 pixels constitute the SW array and the next 37 the LW
array. On the right is a map of the array showing the layout of the pixels. As can be seen there are only
3 pixels above the user-selected threshold of 78 nV. The mean level for the LW array should be around 40
nV, and for the SW array around 60 nV. By design (and some luck...) the most noisy pixels are often to
be found on the edge of the array, which is a far better situation than having one of the more central
pixels affected. (Incidentally, the difference in the mean levels for the two arrays is due to the
passband of the filters - for the SW array the passband is almost twice the LW. The background photon
noise, which dominates the overall level and arises from the cold optics within the cryostat, is
proportional to the square-root of the filter bandwidth, and so the measured noise is higher for the SW
array).
3. What constitutes an acceptable performance?
This is a good question (and may depend on the observation being performed and just how fussy the observer
wants to be...). Since the SCUBA bolometers have a high impedance (some 20 MOhms under optimum bias
settings) they are quite susceptible to changes in the RF environment of the observatory. Associated with
this is the long cable length between the high-impedance bolometer and the low-impedance first stage
amplifiers. This makes the bolometers susceptible to electrical pick-up, changes in the grounding etc. The
upshot of this is that the noise levels can vary substantially on a day-to-day timescale. Some bolometers
(e.g. I9 and G13 shown above) are almost always consistently noisy. Their noise characteristics in the
frequency domain show a distinctive 1/f-type spectrum. This is believed to be caused by a poor contact
somewhere in the signal chain. Over the years we have had some success at repairing these contacts, and
the recent ribbon-cable
upgrade has resulted in a significant reduction (on-average) of the number of noisy pixels.
Pixels exhibiting very high levels can easily dominate the dynamic range of a map, and so in these
circumstances consideration should be given to switching them off for the on-line display. See the SCUBA
Observing Guide for details on how to do this, but note that they continue to record data and can, if
desired, be switched back on for the off-line data reduction. It should also be noted that pixels that are
only moderately noisy (say, a few times the baseline level) can often still be included in the dataset by
using the SETBOLWT feature in SURF. This calculates the noise weights of each bolometer with respect to
the central pixel and can essentially be used to "noise flat-field" your map (although care must be used
in the presence of strong sources). See the SURF manual for more details.
4. Identifying problems
There are quite a few factors that can contribute to excess noise on the SCUBA arrays. These range from
the presence of superfluid helium films to ground loops to low battery supplies. If excess noise is
observed it is generally very good practice to repeat the measurement a few times just to make sure
that a short-lived transient effect (such as a ground spike or cosmic ray...) has not compromised the
data. If the excess noise persists then the following examples can be used to identify the problem and
help to find a solution.
Case 1: Noisy pixels are seen in a small group at the start of the LW array
Symptoms: Perhaps the most common excess noise problem is that associated with contamination caused
by superfluid helium films. This is most likely caused by small amounts of liquid/gaseous helium
penetrating the seal between the LHe can and the vacuum space which houses the arrays. It is believed that
the helium manages to get to the ribbon-cable connectors which carry the bolometer signals to the outside
world. This has been seen on many occasions, and invariably the same set of symptoms are always observed.
The following plots show how to recognise this noise effect when it first appears.
A small group of very noisy pixels is seen at the start of the LW array. The noise level can vary
significantly but is typically many 100's to 1000's of nV. On most occasions the film will move with
gravity - i.e. it starts at pixel G1 (which is closest to the seal) and moves along the connector line.
After a period of time the film moves along the array as shown below (it usually takes about 1 day to move
this amount, although faster changes have been seen in the past).
In this case the noisy pixels would start to have a serious affect on any mapping observation
(particularly for jiggle-map). Photometry using the central pixel would still be possible, although the
loss of 6 or more pixels would start to decrease the effectiveness of the sky-removal algorithm.
Eventually, the situation becomes serious enough that the centre of the array is affected (see below), and
although it is clearly still possible to use the other parts of the array for small, compact source
observations, the performance is seriously degraded and the effectiveness of the array compromised. It
will also be seen that the problem has materialised on the SW array as well. The reason for the "time-lag"
between the first appearance on the LW and subsequently on the SW is not well understood.
Investigation/cure: The standard cure for this problem is to carry-out a "mini-warmup". This
raises the temperature of the array to about 8 K forcing the helium film off the connector pins. The whole
procedure takes about 4 hours, and although there are detailed instructions on how to carry this out, it
should not be done by anyone unfamiliar with the SCUBA cryogenic system. The "mini-warmup" has found to be
effective at temporarily curing this problem for periods of 3-6 days.
In summary:
A mini-warmup should be arranged as soon as the noise moves off the edge of the array and towards the
centre.
Case 2: Groups of up to 16 noisy pixels are observed
Symptoms: When groups of noisy pixels are seen, which are not necessarily at the start of the
array, then it is likely that the problem is not associated with a superfluid film, but most likely is an
electrical problem. A possible exception to this is when SCUBA has not been used for a few days, and
during this downtime the superfluid film has appeared and already moved some way across the array. The
example below shows an excess noise problem that is obviously associated with an electrical problem.
Looking at the SW array it can be seen that almost all the B and F-card bolometers are showing excess
noise, and that the majority of the remainder are okay. The key thing to note here is the 16-channel
grouping. The SCUBA bolometer signals go into groups of 16 at the first-stage amplifier i.e. 16 pixels on
a card (hence the A-I lettering). This problem is most likely caused by a loose connector - probably (by
past experience) on the Tempest analogue electronics box. Other possible causes include a faulty
pre-amplifier card, DAQ card or grounding problem. The situation is not so clear on the LW array in this
case but a similar problem could be in evidence.
Investigation: This can be tricky to diagnose. In the past the 100-way D connector cables coming
out of the Tempest box have occasionally been knocked (e.g. by the LHe dewar during a transfer).
Re-seating the affected connector has cured the problem. On a few occasions the green ground wire (there
is one on each connector shell) has been disturbed. If this is not the cause then it is possible that we
have a faulty preamp or DAQ card. This can be investigated by swapping the suspect card and investigating
whether the excess noise moves with the card change. (Note that some knowledge of the SCUBA electronics is
needed to carry out work of this type i.e. how to use static protection precautions, as well as procedures
for power cycling of various components in the analogue and digital electronics). A less probable
explanation is a fault inside the cryostat (i.e. at the first-stage amps). To fix this would need a full
warmup to ambient temperature, and hence quite serious downtime of the instrument.
In summary check the following in this order:
Check for loose or dislodged cables coming out of the Tempest box (examine and re-seat any that appear
to be suspect)
Check that all the green ground wires are still attached to the same cables.
Swap pre-amp boards in the Tempest rack to see if noise moves with board. A spare card is
available.
Swap DAQ cards as above. A spare DAQ card is also available.
Case 3: High mean noise level observed
Symptoms: In this case the average noise level is much higher than the nominal level. All pixels on
both arays are affected by about the same amount. An example of this is shown below.
The mean noise levels are now about 50% too high for each of the arrays. Although this could be caused by
batteries that are "just-on-the-edge" of being able to supply sufficient power to the FETs, it is more
likely that the background power on the arrays is higher than it should be. Other possibilities include a
low-frequency temperature drift/variation.
Investigation: The first thing to check is whether the reflector blade of the chopper wheel was in
place. Noise levels of this magnitude could be explained by SCUBA viewing the zenith sky and not the
reflector. This means a trip to the Nasymth platform (with a flashlight if it is dark). A large "R"
should be visible on the back of the blade near the window. If the blade is in the wrong position check
the Berger Lahr power supply and cables to the chopper wheel unit. Make sure it has power. If this is not
the problem check for moisture or ice on the window of the cryostat. Again, this could cause an elevation
in the mean noise level (it's a bit like looking at cold or "damp" sky...). A moist window may mean that
the LN gas hose was not re-connected after the daily LN fill. If there is moisture wipe carefully with a
tissue. If there is ice, use a low-powered heat gun to melt (take extra care not to get the window too
hot!). Another possibility that could cause these symptoms is a systematic, low-frequency variation in the
base temperature. Look at the AVS readout and see if there are any low-frequency drifts.
The worst case scenario is that there is "wide-spread" contamination on the arrays (could be
caused by a vacuum leak, for example). In this case a full-warmup will be required to rectify the problem.
In summary:
Check that the reflector blade of the chopper wheel was in place.
Investigate whether there is any moisture or ice on the cryostat window.
Check all batteries.
Check base temperature for drifts or low-frequency variations.
Although unlikely, check for electrical problems e.g. ground loops.
Carry out a mini-warmup.
If all else fails, a full warmup and extended pump-out is recommended...
Case 4: Variable (high) noise levels are seen between successive measurements
Symptoms: In this case successive noise measurements may have little similarity, but in general
will show high, random noise across the array. An example is shown below.
As can be seen there is no obvious pattern to the excess noise, as in the cases of the superfluid He films
or a suspect cable or electronics card. The fact that some pixels are close to the nominal value suggests
that this is NOT a cryogenics problem (e.g. fluctuating fridge temperature). The most common cause of
this noise trace is either an underpowered FET battery or a major electrical ground loop.
Investigation: Check the FET batteries, and also the bias batteries. One problem that has been
encountered in the past has been cabling or metal pipes that cause an additional ground loop. The SCUBA
electrical system has a single-point ground (back at the arrays) and any electrical connections to, for
example, a shell of a cable or a metal pumping line can compromise this. A recent example was that the
spare AVS bridge cable was found to be touching the top plate of the cryostat.
In summary:
Check all batteries (especially the FETs). Replace any that show a low voltage.
Check for problems with the electrical grounding.
If these tests are unsuccessful, check that the fridge is working properly.
Carry out a mini-warmup before anything more drastic.
Case 5: Marginal SW array noise performance
Symptoms: The number of noisy pixels on th SW array is high (say, greater than 20%). If the LW
array is not seriously affected then it is likely that this is caused by an electrical problem associated
with only the SW array.
Investigation: The most likely cause is that the FET battery supply is not supplying enough voltage
to power the SW array first-stage amps. The SW supply usually runs down quicker than the LW supply (simply
because there are more channels to draw current).
In summary:
Check the FET battery supplies. Replace them if low voltage.
Case 6: Marginal SW & LW array noise performance
Symptoms: There are 20% or more excessively noisy pixels on each array. This starts to have a
serious impact on the science achievable with the instrument. An example of this is shown below.
Investigation: There are no obvious patterns in the noise levels in this case. Although the
majority of pixels are close to the nominal baseline level, there are a significant number that do not
meet the specification. The most likely cause of this is low FET or bias batteries, and although the SW
array draws more current than the LW (and should therefore "fail" earlier...) there have been occasions
when both supplies have begun to fail at the same time (battery "fatigue"). In this example the level of
the excess can also give a clue. A ground loop tends to give higher levels (e.g. 100's of nV), and the
superfluid He film symptom even higher (1000's). On two other occasions excessive vibration was found to
be a problem (oscillating antenna drives and excess vibration from the RxB3 compressor). Both these events
caused pixels that have a known microphonic sensitivity to vibration to have higher noise levels. The
worst case scenario is some kind of non-uniform (random) contamination on the array. A mini-warmup is
recommended at this stage, but if unsuccessful a full warmup and extended pump-out would be required to
cure.
In summary:
Check the FET and bias battery supplies.
Check for excessive vibrations from the antenna, CCCs, or SMU.
Carry out a mini-warmup.
As a last resort, carry out a full warmup, extended pump-out and cooldown.
Case 7: Noise levels are too low...
Symptoms This happens maybe a few times a year, and the symptoms are that the noise levels for both
arrays are much lower than the nominal level. Typical levels might be 10 nV or less.
Investigation:Usually this is caused by the FET batteries (i.e. the big rechargeable battery packs)
being replaced but not switched back on. The measurement is then that of the second stage amplifier noise
(about 5-10nV). Another more serious situation may be that the fridge has warmed up, causing the
bolometers to have very low impedance (little sensitivity) to incoming radiation. Contact the designated
SCUBA system engineer as soon as possible if this has occurred.
In summary:
Check that the FET/bias batteries are switched on.
Check the SCUBA base temperature reading.
|