|
Calibration
Next: Continuous Calibration
Up: No Title
Previous: Setting IF Levels
Several options are available.
The recommended option for most accurate calibration and best baselines
is BE DATA continuous (see below). This is also known as
channel-by-channel, as it
uses a separate for every channel in the spectrometer backend.
-
The old methods were FE and IF, where the frontend receiver or IF
power detectors, respectively, are used to give a single
,
etc for each receiver (recently modified to give two of each number, one each
for the signal and image sidebands). This can still be used.
-
One new option is BE POWER ( BE = backend). This produces (and uses)
a separate
, etc for each subband.
-
The other new option is BE DATA. This gives separate values for each
channel of the spectrum. Note that in the case of non continuous
calibration, this should not be done on source
(unless the line is very weak, ie less than
1K), as it would result
in a system temperature which jumps at the line position. For normal
continous calibration, the CAL can be done on-source, as the `SKY' part of
the calibration is not used. It prints on the
operators console an average
value of the , for each subband. Note that each
subband has a number for (the signal sideband) and for
(the image
sideband), using the correct ATM model . The value of may be
much larger than , particularly in the 450--500GHz band, where
the atmospheric transmission is highly frequency dependent. However,
is
the only number that effects the signal:noise ratio of a line in the signal
sideband (ie don't worry too much if is very large).
It is possible to check and display the receiver, sky or system temperatures
across the passband. Load a configuration with NON-continuous CAL, then type:
ICL> send das set das_output ***,
where *** corresponds to TREC, TSKY or TSYS. Then do a
short sample (say 10 seconds in NSW mode). The output is displayed and
stored in a GSD file like a
normal spectrum, as in the example shown in Figure 3. Type
ICL> send das set das_output SPECTRUM
or reload the configuration again to set the output back to regular observing.

The effectiveness of the BE DATA calibration is illustrated in
Figure 4. In the 760-MHz mode the bandwidth covered by
the DAS is wider than the instantaneous bandwidth of the B3i receiver. This
results in a strong increase in the system temperature over part of the band.
However, as shown by a typical spectrum plotted on the same horizontal scale
the only effect is the expected noise increase over the range of enhanced
system temperature.

Next: Continuous Calibration
Up: No Title
Previous: Setting IF Levels
Henry Matthews
Wed May 1 15:19:04 HST 1996
|