Setting IF Levels
Next: Calibration
Up: Loading the DAS
Previous: Loading the DAS
The IF levels on the bar graphs in IF2 unit should be set correctly, and so
the attenuator switches need not be adjusted. With 4 GHz IF
receivers (currently only RxC2), the IF switch internally sets up the
correct power levels every time a CAL is done.
At the end of each sample observation, the Total
Power levels are displayed (one pair for each cycle of a pssw
observation, two pairs for bmsw).
The mean value of these during a sample on the sky should be somewhere
between approximately 3.0 and 8.0; if any of them are above 11.0 or below 1.0,
the IF level settings should be checked accordingly, and the CAL and SAMple
should be redone. Note that if at any time the IF level is
changed, the CAL must be redone.
The DAS can now type out the normalized total power readings of the SKY, HOT
and COLD after each CAL if the PRINT_POWER parameter is set to true.
The default for this parameter is false to prevent extra lines being
sent to the operator's terminal if the information is not desired.
To turn the option on:
ICL> send das set print_power true
and to turn it off again:
ICL> send das set print_power false
Henry Matthews
Wed May 1 15:19:04 HST 1996
|