20061114 report
CW inclinometry too noisy to use
SCUBA-2 preparatory work in the next week or so requires the
removal of one (?) of the basement inclinometers. Hence
inclinometry (CW only) was performed expeditiously as follows:
Datasets El Dirn HST mean leg temperatures Humidity
start start middle end %
( 20061104 89 cw 05:18 - - -
20061104 89 ccw ~08:00 - - - )
20061114 89 cw 05:30 5.6 5.1 4.3
Data is recorded from the 3 active inclinometers only:
there is no strain gauge data, for instance.
-
Preliminary inspection of the data showed increased noise in TY,
as illustrated by the difference plots below:
Click for better image
Normal noise levels in the difference between two sets of TY data (left)
are 7.5mV, whereas the difference between the lastest two TY data sets
(right) has noise (mostly originating from the new data) of 22mV.
A section of the TY difference data is shown below:
It might be fruitful to perform a Fourier analysis on these data,
although one might suspect the elevation servo as the source of
the problem given recent behaviour (see fault [20061024.002]).
(This problem reported as [20061115.006]).
-
The differences between the (CW) model derived from these new data
and the previous model of 01 Nov
are shown below :
There is little systematic change in the track model, but the new data are
too noisy to use.
|