20071223 report
CW & CCW inclinometry OK
Inclinometry was performed as follows:
Datasets El Dirn HST mean leg temperatures Humidity
start start middle end %
( 20071125 88.8 cw ~08:00 2.2 2.7 3.5 20
20071125 88.8 ccw ~10:45 3.5 4.1 4.6 16 )
20071223 89 cw ~07:30 0.7 0.8 1.0 75
20071223 89 ccw ~07:30 1.0 1.6 2.4 75
Data is recorded from the 3 active inclinometers and one of the two
strain gauge channels. The LT trace is abnormally low, so the RT
data is used to correct LX & LY. TY is also +1000mV on average
and a tweak to the elevation used is necessary (web page instructions
amended 20071230).
-
The differences between the (CW) model derived from these new data
and the previous model of 25 Nov
are shown below:
Click for better image
-
The rms scatter in the differences in (F1,F2,F3) are
(0.31", 0.42", 0.41")
cf (0.59", 0.78", 0.65")
previously.
-
The difference between the CW and CCW models is shown below:
. . . and is of the 'small amplitude' type seen regularly this year.
The CW-CCW data were used to generate a
hybrid track model,
installed at ~HST 11:00 on 24 Dec 2007.
The symmetry remains OK;
strain gauge data show a little increase in loads.
|