990111 report
CCW inclinometry
SUMMARY
One CCW inclinometry run was obtained (with the doors closed).
The agreement between the CW and CCW data taken on this day
is better than the agreement between the last such data taken on 980820.
Datasets HST start mean leg temperatures Humidity
start middle end %
( inc990107 (open) 21:15 4.5 4.1 3.7 60
inc990108 (closed) 00:20 3.4 3.1 3.1 65 )
inc990108ccw .. 17:15 4.2 3.9 3.5 80
Shown below left are the CW-CCW differences in the data channels used in
constructing the track model (ie Chs 1, 2, 5, 8 = TX, TY, LY, RY), and
below right similar data from 980820 :
The difference plot shows structures of size up to 2", coincident with
the track segmentation, as has been seen
before .
The forms of the differences are similar but the new data is
generally of lower amplitude. This may also reflect the lack of
symmetry at the earlier
date, as shown in the
symmetry plot .
The rms scatters in each channel confirm this improvement :
| Dataset |
rms Ch 1 = TX |
rms Ch 2 = TY |
rms Ch 5 = LY |
rms Ch 8 = RY |
|
| 980820 |
0.49 |
0.57 |
1.45 |
1.07 |
| 990108
| 0.18 |
0.39 |
1.01 |
0.53 |
Conclusion
- The CW-CCW agreement has improved since the last such
measurements, coincident with the general improvement in symmetry.
Iain Coulson
11 Dec 1999
|