The technical case is an important part of your
application. UKIRT staff use it to check he feasibility of your
observations, to
verify that the time request is adequate and to ensure that we can plan
our instrumentation schedule to meet your requirements.
Source positions
Give as complete a target list as you can - put this
in the technical case if there is not room in the body of the form.
Note that sources above +60 Declination and below -41 declination
cannot
be accessed by UKIRT. Of course you should also
check whether your source RAs are appropriate for the time of year.
Software packages are available to help you schedule your own
observations, such
as UKIRT Rise and OBSERVE
for
STARLINK users.
You must include in your case (estimates of) source
continuum magnitudes or fluxes, or line fluxes if appropriate. This
information is essential for the technical assessor of your programme
to check your quoted exposure times. Please quote these values at the
intended observing wavelength; a V magnitude is of little use to a
technical assessor for a UKIRT proposal. If you
are proposing to observe
a large sample of sources, make sure that you give the mean and range
of
values. Please adopt the following guidelines when quoting the
brightness
of your source.
Imaging
Extended continuum source
Flux density in W/m2/µm/arcsec
or (m)Jy/arcsec2
Point continuum source
Flux density in W/m2/µm,
(m)Jy
or give a JHKLL'M magnitude.
Extended line source
Flux in W/m2/arcsec2
Spectroscopy
Extended continuum source
flux density in W/m2/µm/arcsec2 or (m)Jy/arcsec2
Point continuum source
flux density in W/m2/µm,
(m)Jy
or give a JHKLL'M magnitude.
Extended line source
flux in W/m2/arcsec2
Point source emission line
flux in W/m2
Nb. For line sources, give the line FWHM in µm if it is
resolvable.
Finally, make sure that you use consistent units across
your sample; it has been known in the past for the brightness of four
different objects to be quoted in four different units, which makes
the task of the technical assessor far more onerous than it needs to
be.
Signal-to-noise and required time
You should state your signal-to-noise requirements very
clearly (even if all you say is that you require 2-sigma detections).
Of course, a science case which fails to explain why these are the
requirements is not complete.
The various instrument pages give details of the
exposure times required to reach the background limit, if appropriate.
In this limit your signal to noise will increase as the square root of
the exposure time and in proportion to the source flux. When the
observation is not background limited the S/N increases linearly
with on-chip integration time. This is of relevance for imaging
observations in the
I and Z and, especially, for spectroscopy with UIST and with the
echelle
grating (CGS4) in the J, H and K windows when working between the OH
lines
(echelle background limit reached in around 600seconds).
You should estimate exposure times; do not leave
it to the technical assessor to show that your programme is technically
feasible (or not). It is valid to base your requested time on previous
experience; however if you do this, it is your responsibility to
demonstrate that the objects being observed previously were similar to
those now
being proposed, and to describe the weather conditions on the previous
run.
Finally, if your sources are bright, you should of
course check the saturation limits for your instrument.
Overheads
Allow a proportion of overhead in estimating the total
time required for your observations. This includes peak-up time when
using the CGS4 echelle spectrometer (this is much reduced with UIST
because of the availability of the internal imaging mode), and standard
star observations for all programmes. Consult the web pages for your
instrument for special cases (for example, CGS4
thermal infrared observations, which are dominated by overheads
relating to readout and data transfer time). The following lists
typical values.
Allow 1 minute per object for slews/guide star aquisition.
UFTI read overhead is currently 13 sec/frame for full array. This
adds up if you are asking for 2hrs of on source integration at
60s/frame.
Allow for darks with UFTI
FLATS and arcs for CGS4 take about 3 minutes per wavelength change
CGS4 peak-ups take about 2 min per object (typically; can be much
longer if the source is faint).
Standards (5-10 minutes for a JHK set with UFTI)
UIST imaging acquisition mode is appropriate for targets fainter
than 10th (K). In this case, acquisition for UIST spectroscopy takes of
order 1 minute.
Weather conditions
UKIRT frequently delivers half
arcsecond seeing. Sub-half arcsecond seeing is not uncommon, but
requires flexible scheduling to take advantage of. For this reason
UKIRT is a flexibly-scheduled telescope from semester 03A onwards. You
should specify your programme's seeing requirements in the Technical
Case, even if this is simply to
state that your programme has no particular seeing criteria. Other
weather
criteria include three-micron transparency (function of water vapour
column) and photometric conditions. If you need reasonable transparency
but can do without photometric conditions, you should state this
clearly and explain how your science will be affected.
A description of the statistical behaviour of CSO tau
over the last few years, and a compaison between photometric time and
water vapour, is available at:
The echelle grating covers a very small wavelength
range in one setting. To ensure that the assessor can check your
requirements, be sure to include the resolution and wavelength coverage
demanded by your science, and to determine whether more than one
grating position
is required.
IRCAM
Check filter availability tables, and include filter information in
your case.
UFTI
Check filter availability tables, and include filter information in
your case. For FP work check that appropriate narrow-band filters are
available.
Michelle
Notes on Michelle performance and overheads will be posted here
before the next period of Michelle availability in Semester 04A.
UIST
UIST overheads:
For thermal imaging overheads in array readout can be significant, and
depend
on camera and readout area, see
There are additional observational overheads of target acquisition,
moving
filter wheels and telescope settle time on a jitter - allow an
additional
~20% for these overheads.
Spectroscopy with UIST is more efficient than with CGS4 as there is no
need
to step a grating position to obtain fully sampled spectra, and
acquiring
faint targets can be done by imaging acquisition as opposed to peaking
up
through the slit. A typical flat, arc, standard,target sequence
can be done
with ~80% efficiency.
IRPOL
When using IRPOL with any of the other instruments (Michelle has its
own, built-in waveplate unit), always give the expected polarisation of
the source and the polarisation accuracy needed (see the IRPOL web
pages
for details).
Contact: Andy Adamson. Updated: Wed Dec 1 09:55:00 HST 2004