CGS4 documentation
CGS4 is a 1-5um multi-purpose 2D grating spectrometer containing a
256x256 InSb array, installed in a cryostat which is cooled by liquid
nitrogen and closed cycle coolers. Four gratings are available, two of
which are installed in the cryostat at any one time. They are a 40
l/mm grating which provides resolving powers of 500-2200, a 75 l/mm
grating, a 150 l/mm grating which provides resolving powers of roughly
3000-8500, and a 31 l/mm echelle which provides a resolving power of
about 37,000 (8 km/s). The 75 l/mm grating is rarely used
thesedays. These resolving powers are achieved with the 300 mm focal
length camera optics and a one-pixel-wide slit, which provide a scale
of 0.6 arcsec/pixel for the two moderate resolution gratings and
roughly 0.41x0.90 arcsec/pixel for the echelle (0.41 arcsec/pixel in
the direction of the dispersion).
Introduction to CGS4
General Description of the Instrument
CGS4 is a 1-5um multi-purpose 2D grating spectrometer containing a
256x256 InSb array, installed in a cryostat which is cooled by liquid
nitrogen and closed cycle coolers. Four gratings are available, two of
which are installed in the cryostat at any one time. They are a
40 l/mm grating which provides resolving powers of 300-2000, a 75 l/mm
grating which provides resolving powers of about 600-2000, a 150 l/mm
grating which provides resolving powers of roughly 2000-6000, and a 31
l/mm echelle which provides a resolving power of about 20,000 (15
km/sec). The 75 l/mm grating is rarely used thesedays. These
resolving powers are achieved with the 150 mm focal length camera
optics and a one-pixel-wide slit, which provide a scale of 1.22
arcsec/pixel for the two moderate resolution gratings and roughly
1.0x1.5 arcsec/pixel for the echelle (1.0 arcsec/pixel in the
direction of the dispersion). A 300 mm focal length camera mirror
sometimes is installed in place of the standard camera; with it pixel
scales and wavelength coverages are halved and resolving powers are
doubled (in principle). CGS4 slit lengths are about 80-90
arcseconds. It is possible to orient a slit at any angle on the
sky. Slit widths of one pixel (~1.23 arcsec with the standard camera,
0.61 arcsec with the long focal length camera), two, and four pixels
are available. Since the resolution is matched to one pixel on the
array, with the one-pixel wide slit fully- (or over-) sampled spectra
must be obtained by mechanically shifting the array in the dispersion
direction, while holding the grating fixed. The array may be "shifted"
over two pixel widths; so that isolated bad pixels do not result in
gaps in the spectrum.
For optimum noise performance and efficiency in short exposures
either multiple non-destructive readout of the array (NDSTARE) or a
single read mode (STARE) can be used. The array can also be read in
synch with the chopper. Chopping often is employed when observing at
medium resolution at thermal wavelengths (beyond the K window).
The instrument has a calibration unit, containing a black-body
source for flat-fielding and argon, krypton, and xenon arc lamps for
wavelength calibration. Five broad band filters continuously cover the
wavelength band 0.95-5.4um. Longward of 1.3um, CVFs serve as
order-blockers for the echelle. Shortward of 1.3um, narrow band
filters (1.083um, 1.233um, 1.257um, 1.282um), allow echelle
observations at important wavelengths. Spectropolarimetry is available
with CGS4 at all wavelengths.
Software
The spectrometer is completely under computer control. The
low-level (VAX) side of the control system is run-up and looked after
by the telescope operator, the observer's interaction with the system
is through the ORAC software as with the other UKIRT
instruments. Instrument configurations and observing sequences are
specified using ORAC-OT.
Further Information
Assistance with the CGS4 data acquisition and reduction systems, as well
as other technical information are available on request. Contact persons
are:
- Your support scientist should be the first point of contact.
- Tom Kerr, the CGS4
instrument scientist
CGS4 Optical Parameters
CGS4 optics layout
CGS4 Sensitivity
The 256x256 array has a read noise of about 42e- for 2-read NDR with multi-read
NDR noise of about 23e-.
Sensitivity: Some Very Important Notes - Please Read.
-
The values listed below are surface brightness sensitivities.
-
The sensitivities are calculated for the best part of the band (near 100%
atmospheric transmission). They cannot always be used to interpolate sensitivites
at other wavelengths within the band.
-
Sensitivities for point sources can be estimated from the following
tables. Assuming 0.6" seeing and that the spectrum falls onto three rows
of the array, the signal-to-noise will be approximately a factor 2.5 lower
than quoted in the tables (about one magnitude).
-
CGS4 is less efficient when very bright sources are observed and it is
difficult to obtain signal-to-noise greater than a few hundred. The sensitivities
on very bright sources also do not scale with brightness.
-
In order to achieve the quoted sensitivities, you need to use long exposures.
Typically > 30 seconds in H and K, and >75 seconds at J. Shorter exposures
will result in a lower sensitivity.
Please see CGS4 Exposure Times.
CGS4 sensitivity per pixel with the 40 l/mm grating and the long camera
The 3-sigma 30-minute sensitivities per 0.6"x0.6" pixel for the case of
nodding up and down the 1-pixel wide slit are as given in the following
table. Sensitivities for very extended objects which require nodding to
blank sky will be 0.4 mag poorer. The line flux assumes that the line is
unresolved. Please remember to reduce the sensitivity by one
magnitude for point sources.
| wavelength (µm) |
order |
magnitude |
line flux (W/m**2) |
| 0.9 |
2 |
19.2 |
2.0e-19 |
| 1.25 |
2 |
19.8 |
5.0e-20 |
| 1.6 |
2 |
18.0 |
9.0e-20 |
| 1.6 |
1 |
19.3 |
6.0e-20 |
| 2.2 |
1 |
18.7 |
4.0e-20 |
| 3.4 |
1 |
12.4 |
1.0e-18 |
| 3.8 |
1 |
12.4 |
1.6e-18 |
| 4.9 |
1 |
10.9 |
2.0e-18 |
Predicted CGS4 sensitivity per pixel for the 150l/mm grating and
the long camera.
The 3-sigma 30-minute sensitivities per 0.6"x0.6" pixel for the case of
nodding up and down the 1-pixel wide slit are as given in the following
table. Sensitivities for very extended objects which require nodding to
blank sky will be 0.4 mag poorer. The line flux assumes that the line is
unresolved. In the J-K windows the predicted sensitivity is based on a
model for the dark time inter-OH sky background and the throughput of CGS4.
There will be variation in this background depending on lunar phase. If
you observe an astronomical line which is not resolved from an OH line,
then the limiting magnitide will be less, depending on the strength and
variability of the line. The limiting magnitude is approximately one magnitude
brighter on a typical OH line. Please remember to reduce the sensitivity by one
magnitude for point sources.
Off OH line
| wavelength (µm) |
order |
magnitude |
line flux (W/m**2) |
| 1.25 |
3 |
19.4 |
1.1e-20 |
| 1.65 |
2 |
18.6 |
1.2e-20 |
| 2.2 |
2 |
17.6 |
1.1e-20 |
| 3.8 |
1 |
11.9 |
6.1e-19 |
| 4.9 |
1 |
9.9 |
1.5e-18 |
On OH line
| wavelength (µm)
| order
| magnitude
| line flux (W/m**2)
|
| 1.25 |
3 |
18.3 |
3.0e-20 |
| 1.65 |
2 |
17.0 |
6.1e-20 |
| 2.2 |
2 |
16.5 |
3.2e-20 |
| 2.2 |
1 |
16.5 |
6.4e-20 |
CGS4 sensitivity per pixel with the echelle grating and the long camera
The 3-sigma 30-minute sensitivities per 0.6"x0.6" pixel for the case of
nodding up and down the 1-pixel wide slit are as given in the following
table. Sensitivities for very extended objects which require nodding to
blank sky will be 0.4 mag poorer. The line flux assumes that the line is
unresolved. The values assume that the observations are read noise
limited with sufficiently long exposure times for the multiple NDR to provide
a read noise of about 25e-. The sensitivity will be less if an exposure
time less than 100sec is used to avoid saturating a strong OH line close
to a faint astronomical line. Please remember to reduce the sensitivity by one
magnitude for point sources.
| wavelength (µm) |
magnitude |
line flux (W/m**2) |
| 1.25 |
15.0 |
9.0e-20 |
| 1.6 |
14.6 |
8.0e-20 |
| 2.2 |
14.0 |
6.0e-20 |
| 3.8 |
10.6 |
3.0e-19 |
| 4.9 |
8.9 |
7.0e-19 |
CGS4 Exposure Times
Since the installation of the long focal length camera in August 1997,
many observers have requested that optimum exposure times be placed in
the CGS4 web pages and in the manual. This is not a trivial task, as the
optimum exposure time is dependent on many factors such as resolution,
wavelength, object brightness and weather conditions. In this article,
we provide a guide on how to select appropriate exposure times given the
above factors.
NB. Generally, the maximum possible exposure time is the optimum
exposure time, as overheads are reduced to a minimum. However, you should
discuss this with your support scientist before and during your run.
150 l/mm Grating
Approximate maximum exposure times (sec) for the 150 l/mm grating (J, H, K)
Assume a 1-pixel wide slit and that the light falls on one row of the array.
Typically light falls over three rows and these exposure times can be increased
by about 30%. Half these times when using the 2-pixel wide slit.
|
Magnitude |
| Wavelength (µm) |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
Strong OH1 |
| J (1.2) |
0.16 |
0.40 |
1.0 |
2.4 |
6.0 |
15 |
38 |
95 |
238 |
600 |
| H (1.7) |
sat |
0.14 |
0.36 |
0.90 |
2.3 |
5.8 |
15 |
36 |
90 |
100 |
| K (2.2) |
0.14 |
0.34 |
0.84 |
2.1 |
5.3 |
13 |
33 |
83 |
208 |
120 |
1 The strongest OH line in the band will be saturated using
this exposure time (on a good night).
Approximate maximum exposure times (sec) for the 150 l/mm grating
(L, M)
This table is for 1-pixel wide slit and the long
camera. Half exposures for 2-pixel wide slit.
|
|
FULL ARRAY, NORMAL WELL |
SUBARRAY, DEEP WELL* |
| Wavelength (µm) |
Notes; |
Sky + tel emission; Max exposure (sec) |
0.12s exposure Brightest observable star |
Sky + tel emission Max. exposure (sec) |
0.016s exposure. Brightest observable star |
| 3.0 |
|
20 |
|
40 |
|
| 3.31 |
CH4 Q; |
3 |
|
6 |
|
| 3.2-3.5 |
CH4 v-r lines |
6 |
L ~ 1.5 |
12 |
L ~ -1.5 |
| 3.8 |
|
4 |
L' ~ 1.0 |
8 |
L' ~ -2.0 |
| 4.1 |
|
2 |
|
4 |
|
| 4.7 |
|
0.2 |
M ~ 0.5 |
0.4 |
M ~ -2.5 |
| 5.2 |
|
0.12 |
|
0.25 |
|
* These figures refer to the 256 x 32 subarray. There is also a slightly
larger 256 x 48 subarray with a minimum exposure time of 0.023 sec. Therefore
the brightest observable magnitude is 0.4m fainter when using this array.
150 l/mm grating and background limited exposures.
This medium-to-high resolution grating enables one to work in between OH
lines in many regions in the 1.1 to 2.3 micron spectrum (OH line emission
is not a factor beyond 2.3 microns).
In simple terms, in order to be background limited in the non-thermal
regime (<2.3 microns), the sky noise must be greater than the array
read noise. For multiple non-destructive reads (NDR), the read noise is
approximately 23 electrons. A typical value for the continuum background
(from the telescope, sky and long wavelength leaks) using the 1-pixel wide
slit in J, H and K is 30 counts in 100 seconds, and with a gain of six,
corresponds to 180 electrons and a sky noise of approximately 13.5 electrons.
Therefore, in most cases, an exposure time of approximately 300 seconds
is required for the sky noise between OH lines to equal the read noise.
For longer exposures than this the array is background limited and best
s/n is achieved. With the two pixel wide slit (which still gives high enough
resolution to work between many OH line pairs) the exposure time must be
greater than about 200 seconds. Beyond about 2.2 microns, the background
increases rapidly as the thermal background from the sky and telescope
begin to increase, and the background-limited exposure time drops rapidly.
The drawbacks to using such long exposures are variations in the sky
background and OH line intensities, OH line saturation (at H and K), and
increasing likelihood of spikes on individual or small groups of detectors.
If the critical wavelengths are well clear of the OH lines, then you probably
don't have to worry about their approximate 5-10 minute variation timescales
or strength for OH variations, but if you are close to one than these can
become problems (but see the two paragraphs below for ways to minimise
these). Add to this that you will probably need to oversample your spectra,
your time on source will become at least 600 seconds before you nod to
sky. If you are using the 1-pixel wide slit and 2x2 sampling, it will be
20 minutes before you can nod the telescope. In addition to the dangers
of sky variations these long times mean that, although in principle maximum
sensitivity is achieved, a lot of time is wasted if something goes wrong.
For extended sources, nodding to sky is required, and the brightness
of the source and stability of the sky background on that night will effectively
determine the time between nods and hence the exposure time; these may
be considerably less than the above ideals.
For spectra obtained while nodding along the slit, subtraction of the
negative spectrum from the positive spectrum will remove most of the sky
and OH fluctuations because both vary slowly across the rows of the array.
When observing faint and compact sources it is always advisable to nod
a small number of rows along the slit (e.g., much less than the canonical
30 rows), so that the cancellation of sky and OH residuals is as accurate
as possible. Remaining residuals can be removed by polyfitting techniques,
using blank sky rows adjacent to the rows of interest, but doing this will
increase the noise in the final spectrum.
The frequency of spikes is difficult to judge and their effect difficult
to assess, because spikes sometimes are severe, sometimes are only somewhat
above noise levels, sometimes effect only one pixel, sometimes effect a
few adjacent pixels, and their frequency may vary. Clearly they are more
likely to affect observations of an extended source than a pointlike source.
Empirically they do not appear to be a serious problem when observing point
sources with exposures of a few hundred seconds.
40 l/mm Grating
Approximate maximum exposure times (sec) for the 40 l/mm grating (J, H, K)
Assume a 1-pixel wide slit and that the light falls on one row of the array.
Typically light falls over three rows and these exposure times can be increased
by about 30%. Half these times when using the 2-pixel wide slit.
|
Magnitude |
| Wavelength (µm)
| 3 |
4 |
5 |
6 |
7 |
8 |
9 |
10 |
11 |
Strong OH1 |
| J (1.2) |
sat |
0.4 |
1.0 |
2.5 |
6.3 |
16 |
32 |
64 |
160 |
600 |
| H (1.7) |
sat |
0.15 |
0.4 |
0.90 |
2.3 |
5.8 |
15 |
38 |
95 |
100 |
| K (2.2) |
0.14 |
0.35 |
0.90 |
2.2 |
5.5 |
14 |
35 |
60 |
60 |
120 |
1 The strongest OH line in the band will be saturated using
this exposure time (on a good night).
Approximate maximum exposure times (sec) for the 40 l/mm grating (L, M)
This table is for 1-pixel slit and the long camera. Half exposures for
2-pixel wide slit.
|
|
FULL ARRAY, NORMAL WELL |
SUBARRAY, DEEP WELL* |
| Wavelength (µm) |
Notes |
Sky + tel emission Max exposure (sec) |
0.12s exposure. Brightest observable star |
Sky + tel emission Max. exposure (sec) |
0.016s exposure. Brightest observable star |
| 3.0 |
|
12 |
|
24 |
|
| 3.31 |
CH4 Q |
2 |
|
4 |
|
| 3.2-3.5 |
CH4 v-r lines |
4 |
L ~ 3.0 |
8 |
L ~ 0.0 |
| 3.8 |
|
2.5 |
L' ~ 2.5 |
5 |
L' ~ -0.5 |
| 4.1 |
|
1.2 |
|
2.5 |
|
| 4.7 |
|
0.25 |
M ~ 2.0 |
0.5 |
M ~ -1.0 |
| 5.2 |
|
0.15 |
|
0.3 |
|
* These figures refer to the 256 x 32 subarray. There is also
a slightly larger 256 x 48 subarray with a minimum exposure time of 0.023
sec. Therefore the brightest observable magnitude is 0.4m fainter when
using this array.
40 l/mm grating and background limited exposures.
In most cases the 40 l/mm grating is background limited at much shorter
exposure times at all wavelengths than the 150 l/mm grating due mainly
to its lower resolution, which ensures that an OH line is present in almost
every resolution element.
Typical background-limited exposure times are about 30 seconds at H
and K (less than 2.3um), giving 2 minutes between nods with 2x2 sampling.
In the J band the OH lines are weaker and exposures of ~75 seconds are
required to reach the background limit (5 minutes between nods). The same
concerns and optimal procedures regarding OH fluctuations as discussed
for the 150 l/mm grating apply here, except that spikes are less of a problem
because super-long exposures are not needed.
The Echelle
Optimum exposure times for the echelle are difficult to predict due to
the small wavelength coverage of this grating, the exact wavelength of
a particular observation and the changes in sky transmission with wavelength.
The best course of action is to consult with your support scientist or
with; Paul Hirst before
your CGS4 run.
Resolutions and wavelength coverage (with the long camera)
40 l/mm grating1:
| order |
resolving power |
coverage (um) |
| 2 (0.95 - 1.40 um) |
800 x lambda |
0.32 |
| 1 (1.30 - 5.50 um) |
400 x lambda |
0.64 |
150 l/mm grating1:
| order |
resolving power |
coverage (um) |
| 3 (1.07 - 1.30 um) |
4700 x lambda |
0.055 |
| 2 (1.25 - 2.50 um) |
3100 x lambda |
0.08 |
| 1 (2.10 - 5.50 um) |
1550 x lambda |
0.16 |
Echelle1,2:
- Resolving power is 37000 at the optimum order.
- Coverage is 0.006 x lambda.
1: These are the resolving powers for the 1-pixel wide slit (0.609"). Half the values
when using the 2-pixel wide slit. They will be x4 lower when using the
4-pixel wide slit for extended object, but probably only x2 lower for guided
point sources.
2: There is no 4-pixel wide slit available for use with the echelle.
40l/mm and 150l/mm grating efficiencies
40 l/mm
Use the following plot to decide which order to use when observing with
the 40l/mm grating in CGS4. Use 1st order beyond 2.5 microns.
The data are from Richardson Grating Laboratory for 1st and 2nd order.
3rd order is estimated.
The above plot shows only the grating efficiencies themselves. It does
not include atmospheric transmission and the throughput of the CGS4 filters.
For example, if one wished to observe a feature a 1.35 microns while setting
the central wavelength for the middle of the H band (around 1.6-1.7 microns),
you would expect essentially no throughput at 1.35 microns. This is because
i) the atmospheric transmission itself is only about 20% at this wavelength
and ii) the B2 filter, used for H and K band work, cuts off at about 1.4
microns.
For those wishing to observe features that occur close to or at the
cut-off wavelengths for the near-infrared bands, please contact your support
scientist or Paul Hirst beforehand for advice on the most appropriate
methods.
150 l/mm
CGS4 Filters
The following two tables list the filters in use with CGS4, their
central wavelengths and cut-on and off wavelengths (half power).
Broad Band Filters (for low and medium resolution gratings):
| Filter Name |
Central Wavelength (µm) |
Cut-on Wavelength (µm) |
Cut-off Wavelength (µm) |
Throughput at Central Wavelength (%) |
Estimated Average Throughput1 (%) |
| IJ |
0.94 |
0.84 |
1.04 |
55 |
48 |
| B1 |
1.29 |
0.99 |
1.59 |
83 |
84 |
| B2 |
2.02 |
1.43 |
2.60 |
92 |
89 |
| B3 |
3.15 |
2.14 |
4.16 |
90 |
93 |
| B4 |
3.50 |
2.75 |
4.24 |
90 |
88 |
| B5 |
4.92 |
4.35 |
> 5.5 |
93 |
92 |
| B6 |
2.21 |
2.00 |
2.41 |
78 |
74 |
1 Estimated in the central 75% of the bandpass.
Narrow Band Filters (echelle mode):
| Filter Name |
Central Wavelength (µm) |
Cut-on Wavelength (µm) |
Cut-off Wavelength (µm) |
Throughput at Central Wavelength (%) |
| N1 |
1.089 |
1.078 |
1.099 |
57 |
| N22 |
1.229 |
1.218 |
1.240 |
62 |
| N3 |
1.259 |
1.247 |
1.272 |
62 |
| N4 |
1.275 |
1.264 |
1.286 |
59 |
| N5 |
1.293 |
1.283 |
1.305 |
66 |
2 The N2 filter is not currently installed in CGS4's filter
wheels.
For wavelengths beyond 1.3-µm, the CVFs are used as order-blockers
when using the echelle. When the new wedged CVFs are installed, the narrow
band filters will no longer be necessary, as one of the CVF segments will
extend down to 0.93 um.
CGS4 Pixel Sizes
Low resolution gratings
| Grating |
Perpendicular to slit |
Along slit |
| 40 l/mm |
0.61" |
0.61" |
| 150 l/mm |
0.595" |
0.625" |
Notes:
-
The 40 l/mm grating pixel sizes are independent of wavelength setting.
-
The 150 l/mm grating pixel sizes are virtually independent of wavelength
setting.
Echelle Grating
Because of anamorphic magnification, the Echelle spatial pixel scales vary with grating angle. The 0.41" probably changes with grating angle too, but I don't have measurements of that.
| Grating Angle |
Perpendicular to slit |
Along slit |
| 57.1 deg |
0.41" |
0.783" |
| 62.6 deg |
0.41" |
0.850" |
| 64.4 deg |
0.41" |
0.887" |
| 66.0 deg |
0.41" |
0.93" |
Setting Slit Position Angles (PA)in CGS4
When setting the slit PA in the CGS4 software, it is important to remember:
-
The array is displayed `upside down' on the movie screen
so that south is to the top, and north to the bottom.
-
You need to consider in which direction the nod, or slide, should be specified.
-
Although you are asked to enter a PA which is east of north in your config,
you normally have to enter a negative PA. This is discussed in a little
more detail below.
The best way to demonstrate how to specify the slit PA is by showing some
examples which is done below. In each example, the top and bottom of the
array are marked, as well as north, south, east and west. An arrow indicates
the direction of a positive slide (or nod). Remember the telescope
moves in the opposite direction to this when you slide along the slit,
so be specific when giving instructions to the telescope system specialist.
Please be aware that, in order to try and show things as the observer sees
them on the array when the slit is north-south, north is shown at the bottom
of each plot.
Example 1: PA = 0 degrees (0 degrees east of north).
In this example, a position angle of 0 is entered into the config. the
slit will set to a north-south direction, with the south to the top of
the array. When using a positive slide, the offset beam appears to the
south of the main beam (or towards the top of the array).
Example 2: PA = -45 degrees (135 degrees east of north).
Here, a PA of -45 degrees has been entered. The top of the array is
now SE, the bottom NW.
A positive slide moves the offset beam on the array towards the SE
(i.e., the telescope moves NW)
Example 3: PA = -90 degrees (90 degrees east of north).
A PA of -90 degrees has been entered. The top of the array is in the
east, the bottom is in the west. A positive slide will move the offset
beam to the east, i.e., the telescope moves to the west.
Example 4: PA = -135 degrees (45 degrees east of north).
A PA of -135 degrees is entered. The top of the array is now in the
NE, while the bottom is in the SW. A positive slide will move the offset
beam to the NE, i.e., the telescope moves to the SW.
Some further notes on slit PAs
-
The default slide_slit command in the quad_slide exec is +11.4 arcseconds.
This positions the offset beam 19 rows up from the main beam. To move in
the other direction (towards the bottom of the array), simply enter a negative
value for the slide. Remember that the slit is 90 arcseconds long You normally
do not want to specify a nod which places the offset beam off the array.
-
Although the config demands a position angle east of north, it is only
possible to enter such an angle to a maximum of about 10 degrees. To set
to an angle greater than this, you must specify a negative angle. For example,
if you wish to set to a PA of 50 degrees east of north, set to an angle
of -130 degrees (simply subtract 180 from your desired PA).
-
The pixel size is 0.61 arcseconds. To make sure you nod onto another row,
you need to specify a value in arcesonds which is a multiple of this number.
-
The system usually does a fairly good job of positioning the crosshead
on the slit in the offset position. To increase the accuracy of the
offset position beam, you can do one of two things (or both): i) Use a
shorter slide distance than the default +11.4 arcseconds, or ask the
TSS to peakup in both the main and offset beams and correct the
current slide angle.
-
The current r.m.s. setting accuracy of the slit PA is 0.15 degrees.
And please....
-
Always inform the telescope system specialist when
you change the PA. The peakup position will change with PA, and if the
TSS is unaware of a change, time will be lost in reaquiring your target.
NOTES ON USING THE ECHELLE GRATING IN CGS4.
When using the echelle grating there are more details that need to
be checked than with the low resolution gratings. These are summarised
in this note.
1. Pixel and Slit Sizes.
Due to anamorphic demagnification, CGS4's square pixels do not view
a square patch of sky. Although the solid angle seen by a pixel is
constant, the shape is a function of grating angle. The difference
from a square field of view is very small at the low angles of the
first order gratings, but is quite prominent at echelle angles. For
example, at the nominal blaze angle of 64 degrees pixels are more
elongated in the spatial direction than the dispersion direction by a
factor of ~1.6. With the long camera at angles close to optimum the
pixel size is 0.91 arcsec (spatially) X 0.40 arcsec (spectrally). The
nominally 1-pixel wide slit is in fact about 1.3 pixels wide. The
2-pixel wide slit is 2-pixels wide (ie. 2arcsec spectrally).
The change in shape of the pixels as a function of grating angle is
very rapid at large angles. As a consequence, when using the echelle you
should always measure the size of the pixels by peaking up in two rows.
Edit your sequence to put in the correct offset between the rows
being used when nodding (see also the comment on slit alignment below).
2. Curvature
The slit is slightly curved, so atmospheric and arc lines are not quite
perfectly aligned on one column of the array. In addition the dispersion
axis is also slightly curved, so that the spectrum of a point source is
not perfectly along one row of the detector. Both of these effects are
present with the low resolution gratings, but they are more significant
with the echelle. The curvature of the dispersion direction means that
the dispersion has slight dependance on row on the array. This means that
if an arc/atmospheric/astronomical line is straight along a column of the
array in the middle of the array, lines at the edge of the array will not
be so well aligned. In the dispersion direction the curvature amounts to
approx 0.5pixels across the 256 pixels. In the spatial direction the misaligment
with the columns due to curvature is about 0.1-0.5 pixels close to the
centre of the array, depending on grating angle. For point sources nodded
along about 30 rows the effects are very small and it is still reasonable
to combine the two nodded spectra to cancel sky lines.
There is software available, for example in Figaro, which was designed
to remove these sorts of distortion for optical spectrographs. There is
information about using these in the CGS4 data reduction notes in
correcting
curvature in CGS4 spectra . Using these techniques a residual distortion
of less than 1% can be obtained. However, in conditions of good seeing,
CGS4 data are undersampled in the spatial dimension and this may mean that
the effects cannot be fully removed.
Slit alignment
The curvature of the slits means that even if the postion angle is nominally
N-S on the sky there will be a small (0.3 - 1 arcsec) E-W offset when you
nod along the slit. Since this offset depends on the grating angle and
how far along the slit that you choose to nod, it means that you should
measure it for each echelle wavelength/order that you observe at. After
you have measured the offset between the two rows you need to update your sequence accordingly.
INFORMATION FOR CREATING CONFIGS
3. Optimum Orders
When using the Echelle it is normal to select ECHELLE_AUTO_ORDER as the
grating when defining a config. If this is selected then the CGS4 software
will automatically select the optimum order for your wavelength. The optimum
order gives a higher throughput than other orders. The optimum ordering
software has recently been improved and now works very well for all wavelenghts.
You may wish to use another order, e.g. to get lower or higher spectral
resolution or to increase the spectral coverage. To use a different order
than the optimum one selected automatically, choose ECHELLE as the grating
when defining your config and then explicitely enter the order that you
want.
You can calculate the optimum order for any wavelength as follows. The
blaze angle of the echelle is 64.6 degrees. This corresponds to the product
of wavelength and order (n x lambda) being about 55.0 microns. (e.g., we
find that the transmission at 2.12um in 26th order (n x lambda = 55.1)
is 10X higher than in 25th order (n x lambda = 53.0) In general for any
wavelength, the order for which n x lambda is closest to 55.0 gives the
highest efficiency. However, at lower orders (longer wavelengths), it is
best to fudge this somewhat, as the efficiency drops off more rapidly at
higher angles than at lower angles. The CGS4 software uses a lookup up
table calculated according to the above, with appropriate adjustments at
the longer wavlengths, to select the order for the echelle.
4. Order sorting
Longward of 1.3um, CVFs serve as order-blockers for the echelle. Shortward
of 1.3um, narrow band filters (1.083um, 1.233um, 1.257um, 1.282um), allow
echelle observations at important wavelengths. Wavelengths shortwards
of 1.3um may only be observed through these narrow band filters . The
longest wavelength observable with the echelle is 5.7um. The CGS4 software
will automatically select an appropriate filter or set the CVF to the requested
wavelength. It will report an error if your chosen wavelength is outside
the range of either the narrow band filters or the CVFs. Note that a long
wavelength leak is suspected in the short wavelength CVF (roughly H-band),
and is under investigation. If you want to do Echelle spectroscopy in the
H-band please check with your support scientist for an update on this situation.
5. CVF gradients
There is a slight gradient in the transmission of the CVFs along the slit.
The CVF calibration has been set for the middle of the illuminated area,
row 134. If you want to use rows towards the edge of the illuminated region
or nod more than about 30 pixels you may wish to check the CVF calibration
for the rows you will be using. First of all define an astronomical config
for the echelle in the normal way and set to this config. Peakup a star
on the desired row or look at a lamp line and then run MOVIE. Now while
MOVIE is running go into the menu called DIRECT_MOTOR_CONTROL . This menu
allows you to define an intermediate configuration. The menu items diplayed
represent where the CGS4 motors are currently positioned. To check the
CVF calibration try changing the CVF wavelength by a very small amount
and check whether the signal on the MOVIE display increases or decreases.
Once you know what wavelength you want to set the CVF to calculate the
difference between the grating wavelength and the desired CVF wavelength.
You can then use UKIRT_PREP to save an astronomical config with this CVF
offset. Be very careful if you decide to make a change like this - it is
possible to get "lost" in order on the CVFs because at 1-2.5um the Echelle
orders are very close together. For example 2.2um in 25th order is at the
same grating angle as 2.11um in 26th order - so if you move the CVF by
as much as 0.09um you could be looking at the wrong wavelength and order.
6. CVF fringes
Particularly in the thermal IR a ripple is seen in echelle spectra which
is caused by fringing from the CVF. This ripple can be difficult to remove
if there are amplitude variations between your source and your star. If
you observe very strong ripples, try moving the CVF by a very small amount,
or try puttting your source slightly out of focus. I think the latter helps
because it makes the source and the background fill the slit to the same
degree. Also take oversampled flats in preference to the usual undersampled
one.
7. Wavelength Calibration
Because of the narrow wavelength range covered by the array when the echelle
is used, there are wavelengths where it is impossible to find lamp lines
that fall on the array. In this case you may be able to find lamp lines
at different wavelengths that are present at higher or lower order at the
same echelle setting. I.e., for such a line of wavelength lambda there
is an n that gives the same n*lambda as your observing setup, For example,
you want to observe at 3.00um in 16th order, where there is no line, but
notice that there is an Argon line at 2.40um which in 20th order would
appear at the same echelle angle. Such lamp lines can be found by (1) setting
the echelle to the wavelength you wish to observe (ie in this example 2.40um)
with auto-order selected and (2) changing the arc CVF wavelength to the
wavelength of the calibration line. The arc section of a config allows
you to enter a different CVF wavelength for observing arcs than for observing
your source. This arc CVF wavelength will only be used when you take arcs.
At some other wavelengths you will see lamp lines that were not expected
- these are strong lines in a different order and wavelength being transmitted
through the wings of the CVF transmission profile.
At wavelengths beyond the K window, observable lamp lines are generally
few and far between. For calibration with the echelle, one often must use
the above technique for finding arc lines in different orders, telluric
absorption/emission lines (atlases are available in the control room at
HP, and at JAC), or observations of astronomical line sources.
The Echelle and Peakups etc
There allways seems to be confusion about the Echelle grating and
the need to do 2 row peakups etc. Hopefully, this text outlines the
definative proceedure.
There are two complications with the Echelle that make this more
complex than with the low-res gratings, firstly the curvature of the
slit, and secondly, the fact that the pixel size varies significantly
with echelle grating angle.
Here's my suggested proceedure, assuming that you're nodding along
the slit.:
Initially, set up your ORAC Science Program with the default offsets.
Slew to the target, and ask the TSS to do a 2 row peakup, using
the SAVEPEAK and SAVE_PA commands. You will need to tell them which
two rows you would like them to peak up on. I suggest 140 and 160, the
same as we use for the low res gratings. It will help the TSS if you
can tell them the approximate offset in arc-secs between these two
rows. To do this, look up the grating angle on one of the VAX terminal
screens once you have set to your config, then go to the table in the
"Pixel scales" section of this document and estimate the pixel scale
for the grating angle that you're at.
The SAVE_PA command the TSS uses on the TCS tells the TCS the
correction between the demanded slit angle in your config and the
actual angle on the sky of the point joining the position of the star
when placed on the slit at the two points intersecting the peakup
rows. This means that you can simply ask ORAC to "offset along the
slit" and it will know the exact angle to go to hit the slit on your
two peakup rows.
Note that if you change to a different slit angle, there should
be no need to repeat the two row peakup. Simply do a
normal single row peak up on row 140 - the correction to the angle
will be exactly the same as before, assuming that you're still
offseting the same distance and that you haven't changed the wavelength
or order settings of the grating. If these conditions (ie same
configuration, same size offset) do not apply, then you will have to
repeat the 2 row peakup in your new configuration.
It is advantageous to your signal-to-noise ratio if you have the
spectra centred on pixel rows on the array. Peakup also helps you do
this. After doing the 2 row peakup, the TSS will be able to give you
the exact size of the offset between your two peakup rows. You should
go into the offset iterator in your sequence and replace the default
offset size (usually 11.74) with this number. Again - there is no need
to enter an offset component perpendicualr to the slit unless you
really do want to offset perpendicular to, and hence off, the
slit. This offset size will apply to all observations in the same CGS4
config, no matter what the slit position angle.
Notes on Configuration parameters
GRATING: Select the grating to use. Only two gratings are in
CGS4 at any one time - if in doubt your support scientist or telescope
operator can advise which they are. When observing with the Echelle
you should normally select ECHELLE_AUTO_ORDER.
WAVELENGTH: Enter the central wavelength for your spectrum in
microns.
ORDER: Enter the order to use the grating in. If you have selected
ECHELLE_AUTO_ORDER the value here will be ignored.
The CGS4 sensitivity page contains
information about the appropriate choice of orders for the gratings as
a function of wavelength. Normally echelle_auto_order should be used
for echelle observations. However see the echelle notes for details of where this
is not appropriate. If in doubt your support scientist can advise you.
CVF_OFFSET_(ECH_ONLY): If necessary enter an offset in microns
for setting the CVF when you are using the Echelle. Normally the default
of 0.0 should be used. This value is ignored for the 40 and 150 l/mm gratings.
The precise calibration of the CVF for Echelle observations can depend
on which row you choose to observe, the order and the slit width. Sometimes
it is desirable to set the CVF to a slightly different wavelength from
the grating to correct for this. Your support scientist will help you check
the CVF setting for your wavelength and order. Enter a non-zero value for
a CVF offset with caution and only after checking flats taken with NO CVF
offset.
POLARISER: The polariser should be NONE for normal spectroscopy.
Select PRISM or GRID if you want to do spectropolarimetry.
You cannot do spectropolarimetry with the echelle and prism.
SLIT_WIDTH: Choose a 1-pixel, 2-pixel or 4-pixel wide slit.
There is 1 pixel per resolution element and so a
two pixel wide slit will give reduced, resolution, but can be useful in
poor conditions. The 4-pixel wide slit is not available with the Echelle.
POSITION_ANGLE: Enter the position angle of the slit on the sky
in degrees EAST of North. 0 deg is a N-S alignment. Due to a historcal
reason, CGS4 will not accept angles greater than about 10 degrees. To request
a slit position angle greater than this, you need to specify an angle 180
degrees away. E.g., if you want 45 degrees east of north, you need to specify
-135 degrees as the position angle.
OBJECT_SAMPLING: The sampling required x the number of pixels
to sample over.
There is 1 resolution element per pixel so the detector is stepped to
fully sample. To Nyquist sample your spectra a sampling of 2 should be
selected (ie 2 data points per resolution element). Many users prefer to
slightly over-sample and choose a sampling of 3. To help eliminate bad
pixels from your spectra you can also choose to carry out the sampling
over 2 pixels. This means that each data point in your final spectrum has
been observed with two neighbouring pixels. When the data reduction interleaves
the raw integrations data from a bad pixel is replaced by that from its
good neighbour. Most observers therefore chose sampling of 2x2 or 3x2 for
this reason.
As an example, if you choose sampling of 2x2 the data in the final spectrum
is taken at the following detector positions :
resolution elements | | | | |
detector pos1: 1 2 3 4 5 ...256
detector pos2: 1 2 3 4 5
detector pos3: 1 2 3 4 5
detector pos4: 1 2 3 4 5
Except for at the begining and end, or where there is a bad pixel, each
data point in the final array is the average of two measurements. For 2x2
sampling the reduced spectrum has 514 data points on the x-axis.
DATA_ACQUISITION_MODE: Choose ND_STARE, STARE or CHOP.
For exposure times less than 1 STARE is a little more efficient but
NDSTARE may aso be used. Use ND_STARE for all other observations that do
not require chopping. ND_STARE uses a multiple non-destructive read algorithm
to give optimum read noise performance.
SINGLE_EXPOSURE__SECS: Enter the on-chip exposure time in seconds
to be used for observations of your source. See the optimum
exposure times for a guide to what value to enter here. The minimum
exposure time is 0.12secs for the 256x178 array. For the 256x48 array,
the minimum time is 0.023 secs, and for the 256x32 array, the minimum time
is 0.016 secs.
EXPS/INT_OR_EXPS/CHOPBEAM: Enter the number of exposures to be
coadded into an integration (integ) at each array sample position.
For long exposures on faint sources set object_exp_per_integ to 1. For
very short exposures on a bright source it is a good idea to choose a value
so that the time per integration is a few seconds for best observing efficiency.
CHOP_CYCLES/INTEGRATION This only applies to CHOP mode, and is
ignored otherwise. Enter the number of chop cycles you want before nodding
the telescope (one chop cycles is an object-sky pair).
FLAT_SAMPLING: choose AS_OBJECT in most cases
For Echelle users it is now possible for the data acquisition to take
over-sampled flats with the same sampling factor as your object observations
to help remove CVF ripple.
FLAT_LAMP: choose your required lamp. The numbers refer to black-body
apertures. There is a table of flats
of appropriate values as a function of grating wavelength and order.
If you choose off then the calibration unit will not be used - useful for
if you want to observe the sky, or the dome etc for a flat.
FLAT_ACQUISITION_MODE: Normally this will be NDSTARE
FLAT_EXPOSURE_SECS: typically 0.12 - 1 sec. See the
table of flat exposure
times.
FLAT_EXP_PER_INTEG: The number of exposures to be coadded into
an integration. typically 50-100.
FLAT_INTEG_PER_OBS: The number of integrations to be taken and
returned to the Vax. Typically 1.
ARC_LAMP: Choose your required lamp. There is a table
of recommended arcs lamps as a function of wavelength. Plots of sample
arc spectra can also be found at that page. If you choose OFF for the lamp
then the calibration unit will not be used - useful for if you want to
observe sky lines for wavelength calibration.
ECHELLE_ARC_CVF_WAVELENGTH: Defaults to the same wavelength as
for your object observation.
Because the wavelength coverage on the array with the Echelle is not
very large it is possible to observe at wavelengths where there are no
lamp lines which would fall on the array. This allows you to select a different
CVF wavelength from the grating, to look at lamp lines in a different order
from your source.
If the value shown here is -1 the default CVF setting - ie the same
CVF wavelength as for your object will be used. If you have just recalled
or created a CONFIG then the default wavelength should appear here.
ARC_ACQUISITION_MODE: Usually NDSTARE
ARC_EXPOSURE_SECS: Typically 0.12 secs with the 40 l/mm grating
and long camera.
ARC_EXP_PER_INTEG The number of exposures to be coadded by the
into an integration at each detector sampling position. Typically 5.
All ARCS are automatically taken with the same sampling and sampling
range as your object observations. Although on-chip exposure times are
short, you should add sufficient exposures so that the time per sampling
position is at least 0.5 secs, to prevent the background increasing due
to the detector translation stage getting warmer.
ARC_INTEG_PER_OBS The number of integrations to be taken and
returned to the Vax at each detector sampling position. typically 1.
CGS4 Latency
It has come to our knowledge that the 256 x 256 InSb array in
CGS4 suffers from small scale latent images. This problem has only
become noticeable since long integrations have become possible due to the
small pixel sizes in use with the long focal length camera.
In general, the latency only affects long exposures of very faint sources
after previously observing a bright source (e.g., a standard star). The
level of latency varies, but is usually less than about 5 counts (30 electrons)
above the general background level.
The simplest method to rid the array of the latent image is to ask
your TSS to run manpeak to take a few short exposures without changing
the apperture, once you are pointing at your faint source.
If your targets are bright enough to peakup on, then the above procedure
is unnecessary. The process of peaking up will remove any latent image.
CGS4 Object acquisition Procedures
Peak-up procedures with CGS4 seem to be a large source of
confusion, especially for observers determining before hand what their
requirements are in terms of astrometry, co-ordinates and reference
stars. This page aims to clear things up a bit.
The complication arises in that there are many ways to attempt to
ensure that you have accurately hit your target with the slit, all of
which apply to different situations.
Concepts and Introduction
The aperture defines the position of the slit within the
telescope focal plane. The default aperture is re-determined during
engineering time whenever the instrument has been off the telescope,
and at regular intervals inbetween. The aperture varies slightly with
flexure etc. Fine adjustments to the aperture definition are made
with the peak-up procedure.
The actual peak-up procedure is the process whereby the
aperture definition is finely adjusted to give the maximum signal on
the CGS4 array, whilst pointing at some (bright) target. This
involves taking several (typically 15 or so) CGS4 integrations. Each
of these must convincingly detect the object being peaked up on. It is
possible to automatically subtract sky frames from the images used
for peak-up. The integration time needed depends on the instrument
configuration in use and on how bright your target is.
With the 40l/mm grating, peaking up on a target fainter than 14th
magnitude takes a long time. If you need to estimate peak-up
times for fainter targets or other configurations, refer to the CGS4
sensitivity page, and use the following: Calculate the exposure time
for a 3 sigma detection of a point source at your target's
magnitude. Multiply this by a factor 2 (to account for the fact that
for most of the frames, you're not peaked up, so you're only seeing a
fraction of the light). Multiply this by 15 (say 15 observations
needed).
The Fast Guider is mounted on the cross-head. The
cross-head can position the fast guider very accurately (~0.1") within
the telescope focal plane. The crosshead limit is roughly a circle,
180" from the pointing centre (ie centre of the slit).
The Fast Guider brightness thresholds also vary with
conditions. Because the fast guider works in the optical is is
affected by scattered moon-light. In exceptional conditions -
clear skies, exceptionally good seeing and no moon, we can guide on
stars down to V magnitude of 18.7. A V mag limit of 16 or 17 is more
realistic for average conditions. On fainter targets, we increase the
integration time of the autoguider CCD, which reduces the Tip-Tilt
frequency. On bright targets, we guide at 100Hz. We can go down to
~20Hz for faint sources. Bright sources are more likely to give you
really good Tip-Tilt performance.
UKIRT employs a dichroic tertiary mirror, feeding the IR light to the
instrumentation, and the optical light to the fast guider. This means
that it is possible for the fast-guider and the IR instrument (eg
CGS4) to observe the same target.
Trivial case - bright (~14th Mag or brighter in the IR) guidable targets.
In this case you guide on the target, and you peak up on it before you
start taking data. The peak-up takes a few minutes.
No special co-ordinate, astrometry or guide star requirements.
Bright (14th Mag or brighter in the IR) non-guidable targets.
Reasons they might be non-guidable include: Target not point-like in
the optical, or target is so red that despite being bright in the IR,
it's too faint for the fast guider to see (see notes above).
In this case, you guide on an offset guide star (needs to be guidable
and within 180" of target), and you peak up on the target.
Requirements: A guide star. You can pick one using the ORAC-OT
preparation tool when you get to Hilo. You don't need any special
astrometry. If your co-ordinates are off by more than an arc-sec or
two, peak-up will take longer.
Intermediate brightness targets.
OK, the definition of "intermediate brightness" is non trivial. By
this, I mean something you can guide on, but something that
would take too long to peak up on. See the notes above about the fast
guider sensitivity and peak-up times.
Procedure: The TSS will slew the telescope to a nearby (within a few
degrees) bright CMC star, and will peak up on that whilst guiding on
it. This ensures that the slit and fast guider are at the same place
within the telescope focal plane. Then they'll slew back to your
target and guide on it. Because peaking up on the CMC star aligned
the guider and the slit, the slit will now be accurately placed on the
target.
Note: The guider and slit will not move relative to each other over
a small slew to the CMC and back. If it were a long slew, there would
be concerns about flexure and differential refraction between the
optical and IR beams.
Requirements: The TSS will locate a suitable CMC star as needed. You
need reasonably accurate co-ordinates (ie within a few arc-sec). If the
target is towards the fainter edge of the guider's capability, better
co-ordinates will help. If the source is not point-like enough to guide
on, treat this as a faint target.
Faint targets.
By "Faint", I mean too faint to guide on. This obviously implies also
too faint to peak up on. There are several options in this case.
Blind offset
This is the most reliable way of acquiring very faint targets. You
need an accurate astrometric offset between your target and a
nearby (within 180") guide star (see notes above on fast guider for
what qualifies as a guide star). You would usually measure such
offsets from a deep image of your target field. By "accurate", I mean
to within a fraction of the slit width you will be using. Say
0.2" for the 0.6" slit. The more accurate your offset, the more light
you'll get down the slit.
Proceedure: Simply enter the RA and Dec co-ordinates for the target
and guide star in the OT. Ensure that the co-ordinates you enter have
the correct offset between them, taking into account the factor 15
between seconds of RA time and seconds of arc. When you slew to the
field, inform the TSS that you have an accurate crosshead offset set,
and thus to adjust the telescope position rather than the crosshead
position to bring the guide star into the fast guider if necessary.
The TSS will ensure that the telescope pointing model is good prior to
you starting to take data. This will involve at least going to a
nearby CMC star, and probably doing a peak-up on it. If necessary they
will collimate the telescope pointing model to the CMC star
co-ordinates at this point.
Requirements: Accurate astrometric offset to a nearby guide star.
Blind Pointing
This is the "last resort" method, though the UKIRT Telescope Control
System and Pointing model is sufficiently good that this method
usually works well.
You would use this method if you don't have the astrometric offset to
a guide star necessary to carry out the previous method. You do,
however, need accurate absolute co-ordinates. You shouldn't rely on
this method with the 0.6" slit. Many observers have had great success
using this method to observe radio sources, with accurate co-ordinates
from high resolution radio interferometers, using the 1.2" slit.
Proceedure: Type your accurate co-ordinates into the OT. Use the OT to
find a nearby offset guide star. The TSS will go to a CMC star close
to your field and collimate the telescope pointing model. When they
slew back to your field, inform the TSS that the target co-ordinates
are accurate, and to move the crosshead to bring the guide star into
the guider if necessary.
Threats: You need accurate, co-ordinates. You might need to be
aware of the co-ordinate frame your co-ordinates are referenced to
(how well is this frame tied to the Hipparcos or Radio reference
frames?). This method becomes less reliable in high wind conditions -
during the time the TSS is adjusting the guider position to acquire the
guide star, the telescope is sat on it's main drives, as opposed to
being locked onto a guide star. If wind gusts are knocking the
telescope around at this point, the drives are unlikely to be able to
hold it in position to great accuracy.
Requirements: Accurate absolute co-ordinates.
Absolute Desperation
If your targets are too faint to guide on, and you don't have
either sufficiently accurate co-ordinates or astrometric offsets, then
your best option is probably to use some time at the start of your
observing run to get UFTI images of your fields, from which you can
measure offsets in order to do a blind offset. You should discuss this
with your support scientist and Paul Hirst, the CGS4 instrument
scientist, well in advance of your arrival in Hilo.
Preparing for CGS4 observations with the ORAC-OT
This document overviews the preperation phase of observing with
CGS4 under ORAC. This mainly involves the use of the ORAC-OT - the
Observing tool. Whilst you will no doubt be running the OT whilst you
are observing to make any last minute changes, you should have the
bulk of your preperation done before arriving at the telescope.
General ORAC-OT introduction
The general
ORAC-OT documentation overviews the various concepts of the ORAC
system, and makes a good introduction to this more specific document
on how to use the ORAC-OT to prepare for CGS4 observing.
Getting started with the ORAC-OT
First up, launch the oracot and open up the "UKIRT Template
Library" from the file menu. Also in the File menu, select "new
program" to create yourself an empty program.
Now a trivial example of editing soemthing: In the new program
window that just appeared, double click on where it says "Science
Program". An edit window appears immediately to the right of your
science program window. Type in a title for this science program into
the box labeled "Title". Ignore the other fields.
Basically, you prepare your science program by copying items from
the template library, and editing them to your needs. In the template
library, open the CGS4 folder by clicking on the little handle icon to
it's left - the handle twists down and you can see the contents of the
section. You should see a list of template cgs4 observations, looking
somewhat like this:

Now, you're going to want to do the array tests at the start of your
observing, so copy the "array tests" observation into your science
program. To do this, put the mouse over the array tests, press and
hold the left mouse button and move the mouse into your science
program window. Now the OT is a bit picky about exactly where you drop
stuff. Move the mouse (still dragging the little blue square
observation icon") over the title or icon of your science program. As
you go over this, you'll notice a little red and white diagonal arrow
sign appear over your science program icon. This indicates that if you
drop it there, the array tests observation will go into your science
program, which is what you want, so go ahead and release the mouse
button. You should now have something like this:

Adding a Flat and Arc
Next, you will want to take a flat field and arc lamp
observation. Drag over the "flat and an arc" observation from the
template library. This time, when you drop it, you'll want to have the
pointer over your array tests, and you should see a white and red
arrow sign, with the arrow pointing straight down.
A note on the little arrow signs: The arrow pointing straight down
indicates that when dropped, the new item will be placed below the one
where the arrow appears. A diagnoally down/right arrow indicates that
the idem will appear inside the one where the arrow appears. We'll see
more of this later.
At this point, you might like to change the title of the flat and
arc, for example to add a comment about which band this is. Do this in
the same way that you edited the title for your science progam.
Now, we'll look inside the flat and arc at the components that make
it up. Click in the little handle next to "flat and an arc" to expand
the view. Notice how things inside other things get indented to the
right (inside the flat and arc observation are four components, one of
which is a "Sequence" component, itself containing more
items). Whenever something contains things inside it, it has a little
handle sign to it's left, demoting whether it's being displayed
expanded (handle vertical) or condensed (handle horizontal). Simply
click the handle to toggle this. Things should now look like this:

The "CGS4" component contains the instrument configuration for
CGS4. Double click on it to bring up the CGS4 configuration
editor into the right hand panel of the window:

Edit any of the values as necessary. Don't worry about the exposure
time and coadds that you set here, as these ones aren't actually used
- the exposure times for the flat and arc are set specifically, as
we'll get to shortly. When you're happy with the values, click on the
"Flat" component inside the sequence. You should get a flat component
editor looking like this:

Note carefully: The parameters for the flat field
observation don't automatically update if you edit the fields in the
CGS4 configuration component. To have the software estimate the flat
field parameters based on what's set in the CGS4 configuration
component, click the "Default" button in the flat field editor. If you
then go and change the settings in the CGS4 component, you'll need to
go back into the flat field and hit default again. If after taking a
flat, you find that the default values aren't quite right, then you
can simply edit the exposure time etc directly in the flat field
component.
Now go into the Arc component, and hit the default button to get
the default arc parameters for the configuration you've selected.
Adding a Standard Star
We have a library of bright G stars built availiable in the OT to
speed things up here. Don't worry if you want to use a star that's not
in there, we'll get back to that shortly.
Open up the CGS4_bright_G_library from the file menu of the main OT
window. Simply pick a star, then drag it into your science program,
taking care when you drop it that you've got a vertically downward
pointing red and white arrow sign on the flat and arc observation. You
might want to collapse the display of the flat and arc first, by
clicking the little minus sign to the left of the "flat and an arc"
label.
Once you've got a star there, you need to edit that observation to
match the configuration you're actually using. There's an easy way to
do this. First expand the flat and arc observation and the standard
star observation so that you can see all their components. You don't
need the details of the sequence at this point, so you can collapse
the sequences in each to save screen space if you like. Now double
click the cgs4 component in the standard star, and note the magnitude
of the star. Now, single-click the CGS4 component in the flat and arc
observation, and click the copy button (the one on the right of the
scissors (cut) button). Now single click the CGS4 component in the
standard star, and hit the paste button (the one to the right of the
copy button - looks like a clipboard). When it asks if you want to
replace the component, say yes. Now double click the CGS4 component in
the standard star to get the editor window back, and set the magnitude
back to what it was. The software will then autmatically set the
exposure time and co-adds for that magnitude. you can over-ride these
manually by simply changing the values. We're still working on the
defaults so you might have to do this. The best thing to do is to ask
your TO for the count rate after they complete the peakup, then set the
time to give you around 3000 counts.
Adding a Target observation
First of all, you have to decide which of the observations in the
template library is most appropriate for your target. For point
sources, this is probably point source, nod along slit. For
extended sources, you'll want extended source, nod to blank
sky. If you are in doubt as to which sequence to use, ask your
support scientist or contact the CGS4 instrument scientist for
advice.Drag the one you want from the template library into your
science program. As an example here, we'll use the point source,
nod along slit one. Drop it so as to appear after your standard
star.
Now, in your science program, double click on where it says "point
source, not along slit" and in the editor panel that pops up, type in
a title for this observation. This isn't used as the target name or
anything, it's for your own reference, and should be reasonaby
descriptive as you will choose this name from a list of your
observations to specify that this is the one you want to observe.
Setting the target co-ordinates etc
Now click the little handle next to your target observation to
expand the view, and click on the "Target List" component. In the
panel that appears, fill in your target name (this is what will go
into the data headers etc) and the RA and Dec of your target. If you
wish, you can enter a target name and click the "Resolve" button to
have the OT query the target name on SIMBAD and get the co-ordinates
from there. If you want the name resolver to query an different
database, simply select it in the pull down to the left of the resolve
button.
Next, we'll go into the plot window to check things and to set a
guide star etc. Click on the "Plot" button in the lower left of the
target editor window. In the window that appears, go into the
"Calalog" menu and under "Image Servers", select "Digitised Sky at
JAC". After a pause and some flashing lights, a digitised sky survey
image of the sky area you're looking at should fill in. To the lower
left of the image are a set of togle buttons, collectively labeled
"View". Turn on the ones for "Sci Area" - shows the slit, "Dichroic"
and "X-head" - the latter two between them show you the area of sky
from which you can select a guide star. Now go to the "Catalog" menu
again and under "Catalogs", select "Guide star catalog at CADC". When
the progres box disapears, a list of HST guide stars will have been
loaded. Turn on the "Catalog" view button, and these will be
highlighted on the sky survey image. To manually inspect the list of
stars, select "Browse" from the "Catalog" menu.
To the left of the window, there's a selection of buttons labeled
"Mode". Click on the one labeled "GUIDE", now point and click on the
star that you wish to use as a guide star. If you select one of the
ones from the catalog, the positions from the calalog will be used. If
you prefer, you can just click on a star at any position on the sky
survey image. If you're doing this, then you might want to use the
zoom facility to give you more accuracy. You must select a star that
is within the purple X-head circles (and avoid ones close to the edge
as you're going to be nodding) and not hidden by the green dichroic
edges (again, and not too close).The position editor window should now
look something like this:

Returning to the target editor panel in your science program
window, you'll notice that it has filled in the guide star
co-ordinates you chose, and if you edited your position using the
"drag" facility, the co-ordinates have updated.

Setting Exposure times etc
Next, go back to the flat and arc observation and copy the CGS4
component again. Replace the one in your target observation with this,
then go into it and set the magnitude or exposure time that you wish
to use.
Note: The BL option in the magnitude pull down selects the minimum
time necessary to become background limited. You might want to observe
for longer than this, but bear in mind that the time between nods
equalls this time multiplied by the coadds and the sampling. See the
CGS4 handbook for more details.
The sequence
The only thing you'll usually wish to change in the sequence is the
number of repeats. The "Offset" itterator executes a 4 position jitter
(Object - Sky - Sky - Object). The repeat iterator sets how many times
to repeat this sequence. Simply double click the repeat iterator and
set the number.
Data Reduction Recipies for CGS4
In general, if you based your observations on the template library,
the DR-RECIPE component will be all set up for you, and you won't need
to chnage anything.
The ORAC-DR section of the CGS4 manual provides more info on what
the individual recipies do and the observing sequences they are
appropriate for. If you are unsure or if you think you will need a
recipe that is not provided in our standard selection, you should
contact your support scientist or the CGS4 instrument scientist
well in advance of your observing run.
The Mauna Kea OH Line Emission Spectrum
The following spectra show the OH line emission spectrum in the J, H, K
windows at UKIRT. The spectra were taken with CGS4 using the 150 l/mm
grating at 3rd order in J, and 2nd order at H & K and the 256 x 256
InSb array (0.61" x 0.61" pixels).
At the bottom of the page are links to more
detailed spectra of individual regions of each band. Use these to check
the details of the OH emission lines in your wavelength region of interest.
The spectra were wavelength calibrated using this table of OH
wavelengths.
-
Please be aware that each spectrum is not shown to the same scale. The
OH lines in the H window are much stronger than in K, and those in K are
stronger than those in the J band.
-
The flux units are given as counts, 1 count = 6 electrons, and are per
pixel per 100 seconds.
-
The spectra are not corrected for atmospheric transmission.
-
Wavelengths are in vacuo.
-
The solid angle for a 0.61" x 0.61" pixel is 8.7455x10-12
sr
-
The data were obtained by Tom Geballe and Tom Kerr on UT date 1997 19 December
-
The CSO taumeter
read approximately 0.05 while the data were obtained. This corresponds
to an H2O column of 1mm, a good night on
Mauna Kea.
Click on any of the three spectra below to see a larger scale plot of the
same data.
J Band
H Band
K Band
Individual wavelength regions
The spectra below show individual spectral segments covered by the 150
l/mm grating at the listed wavelengths. Approximate airmass is given in
brackets. In the J band 3rd order was used (wavelength coverage 0.055um),
2nd order was used for H & K (0.08um).
J Band
1.14microns (1.23),
1.18microns (1.24),
1.22microns (1.25),
1.26microns (1.26),
1.30microns (1.27),
1.34microns (1.29)
H band
1.40microns (1.30),
1.46microns (1.23),
1.52microns (1.23),
1.58microns (1.24),
1.64microns (1.24),
1.70microns (1.24),
1.76microns (1.24),
1.82microns (1.24)
K band
1.88microns (1.25),
1.94microns (1.26),
2.00microns (1.26),
2.06microns (1.26),
2.12microns (1.27),
2.18microns (1.28),
2.24microns (1.29)
Atmospheric Transmission
See the details in the UKIRT astronomy calibration area,
here.
Atmospheric Extinction
See the information in the UKIRT astronomy calibration area,
here
Suggested Flat Exposure Times
40 l/mm grating:
| Wavelength |
Order |
BB Aperture |
Exposure Tim (secs) |
| 1.15 |
2 |
5.0 |
2.0 |
| 1.30 |
2 |
5.0 |
2.0 |
| 1.70 |
1 |
1.3 |
0.6 |
| 2.20 |
1 |
1.3 |
0.25 |
| 3.10 |
1 |
3.2 |
0.75 |
| 3.80 |
1 |
2.0 |
2.0 |
| 4.90 |
1 |
1.3 |
1.0 |
150 l/mm grating:
| Wavelength |
Order |
BB Aperture |
Exposure Time (secs) |
| 1.15 |
3 |
5.0 |
5.0 |
| 1.30 |
3 |
5.0 |
2.5 |
| 1.70 |
2 |
2.0 |
1.0 |
| 2.20 |
2 |
3.2 |
0.25 |
| 2.20 |
1 |
3.2 |
0.12 |
Echelle:
Because of the extremely small wavelength coverage of the echelle, it is
fairly impracticable to provide a table of suggested blackbody apertures
and exposure times. The quickets way to determine an appropriate exposure
time and aperture is as follows (this technique can be used with any grating):
-
Write an astronomical config for the wavelength you wish to observe at.
-
Set to this config using the Set_Single_Observation menu in the CGS4 software
and specify a flat observation.
-
Select "Movie", select an exposure time in "Set_Movie" (e.g., 0.5 secs)
and then "Run_Movie"
-
Note the counts on the movie screen.
-
Select "Direct_Motor_Control" and select another blackbody aperture. You
will see the counts change.
-
When you are happy with your selection (i.e., 3000 - 4000 counts) abort
the movie and edit your config with the appropriate numbers.
CGS4 Arcs
For a given wavelength, we do recommend that specific arc lamps are used.
These will give the best coverage of unsaturated spectral lines. They are:
| Wavelength |
Order |
Recommended Lamp |
Alternative Lamp |
| 1.15 |
2 |
Krypton |
Xenon |
| 1.30 |
1 |
Krypton |
Xenon |
| 1.70 |
1 |
Xenon |
Argon |
| 2.20 |
1 |
Argon |
Xenon |
| 3.10 |
1 |
Xenon (2nd order) |
Argon (2nd order) |
| 3.80 |
1 |
Argon (2nd order) |
Xenon (2nd order) |
List of available arc spectra.
Below is a list of arc spectra obtained with CGS4 at UKIRT. You may use
these to help wavelength calibrate your CGS4 data both during and after
your run. Currently, only low resolution spectra (75 l/mm grating, short
focal length camera) are available on the web.
Lamp lines in the 3 to 5 micron region cannot be observed due to absorption
by the lamp bulb glass. However, you can obtain lines in this region by
observing 1 and 2 micron arc lines in 2nd order. In the spectra listed
below, we have not indicated the wavelength value in 1st order for lines
beyond 3 microns. If you are interested in this number, simply divide the
indicated value by 2.
GIF files:
Argon
1.15um,
1.30um,
1.60um,
2.18um,
3.10um,
3.40um,
3.80um,
4.70um
Krypton
1.15um,
1.30um,
3.10um,
3.40um,
3.80um
Xenon
1.15um,
1.30um,
1.60um,
2.18um,
3.10um,
3.40um,
3.80um
PostScript files:
Shift-Click these to save them to local disk.
Argon
1.15um,
1.30um,
1.60um,
2.18um,
3.10um,
3.40um,
3.80um,
4.70um
Krypton
1.15um,
1.30um,
3.10um,
3.40um,
3.80um
Xenon
1.15um,
1.30um,
1.60um,
2.18um,
3.10um,
3.40um,
3.80um
Observing with CGS4 and ORACDR
This document gives a guide to the oracdr pipeline for CGS4 data. It
will give you a preview of how your data will be reduced and displayed
whilst you are at the telescope observing. It might also be usefull
for anyone running oracdr on CGS4 data offline.
General ORAC-DR documentation
The ORAC team have produced some
general ORAC-DR
documentation, which overviews the software. The rest of this
document gives information specific to using ORAC-DR with CGS4.
Running up the pipeline
Type runup at a shell prompt to get current instructions on
starting the pipeline.
Other things to note: If for some reason you need to stop the pipeline
at any point, simlpy issue a Ctrl-C in the xterm window where
you started it. If the pipeline is part way through a recipe at the
time, several Ctrl-Cs might be required, and there will be a
short delay before the pipeline exits.
If you suspect that the pipeline did not exit cleanly, issue the
oracdr_nuke command to tidy up any crash debris.
To re-start the pipeline, use: oracdr -loop flag -from
NUMBER, where NUMBER is the frame that you want your
re-started pipeline to start from. You should generally re-start the
pipeline from the first observation in a group. If you're not sure
where the group starts, the oracom window shows you the current group
number, which is equal to the number of the first frame in the
group. If you're a group or more ahead of yourself, then see the later
section on viewing or re-creating a nightlog
file.
Display Overview
You can change what the
pipeline
display system shows you and how it displays it by editing the
disp.dat file, either directly or by using the
oracdisp
utility.
The default displays should be fine for most people. These are
described here. The main reason why you'd want to change anything from
the defaults at the moment is that when plotting a spectrum, the display
system will autoscale, which you don't want to do if there's a large spike
or other anomoly in the data. See "Changing the plotting scales" below for
more details.
Generally, two windows are used to show you data as soon as possible
after it comes off the array; one GAIA window, which shows you
an image containg the data, and a KAPPA view window which shows
histograms, extracted spectra and the like.
A similar two windows are used to show you more reduced data to
display intermediate results and show you how well you're accumulating
your signal to noise ratio etc. A GAIA window shows you the
current group file image - ie the sky subtracted frame, or the
difference of the main beam and offset pairs aquired so far. A
KAPPA window is then used to show you the divided by standard
and flux calibrated spectra.
Changing the plotting scales
The pipeline display system is controlled by a text file called
disp.dat, which resides in your $ORAC_DATA_OUT directory.
You can edit this file manually with your favourite text editor, or you can
use the oracdisp utility.
The disp.dat file contains one line for each type of data that is to be
displayed, specifying the file type that this refers to, the plotting tool to
use and various parameters of the plot - ie axis scales, autoscaling etc.
One important thing to remember is that the display system allways works in
array co-ordinates, ie. x is wavelength, y is along
the slit, and z is the data value.
To turn off autoscaling of your spectral plots:
- Go either to the xterm where you launched oracdr from, or get a new
xterm and run oracdr_cgs4 in it.
- run oracdisp
- Select the line for the type of plot you wish to change. This will
probably be DBS for the divided by standard spectra, or FC
for the flux calibrated spectra. Double-Click on the appropriate line in the
lower pane of the oracdisp window. You may have to scroll down using the
scrollbar to find the line you want.
- In the upper portion of the oracdisp window, find the Z: line
and click the button next to "Set" to go from autoscale to
set mode, and change the zmin and zmax values to
the axes range you would like displayed.
- Click the "Modify" button at the bottom of the top pane of the oracdisp
window.
- Click the "Configure" button at the bottom of the oracdisp window.
The new settings will take effect next time a spectrum of that type is
displayed, which if you're still oberving the same group will be when the next
pair of observations is complete.
Starting Observing
Array Tests
The first data you should take are the ARRAY_TESTS data. When the
first of these frames arrives, two of the aforementioned display
windows will be opened to display them.
When the array tests have completed, oracdr will analyse them and
print out a message telling you whether the array performance is as
expected. The actuall text should end with something like this:
Double correlated readnoise = 40.1 electrons is nominal
Median Dark current = 0.89 is nominal
Modal Dark current = 0.44 is nominal
If it declares any of the values to by HIGH, then you should consult
your support scientist, or the CGS4 instrument scientist, for advice.
LOW values are beneficial, not a problem!
Flat and Arc
After completing the array tests, you will probably want to take a
flat field and an arc spectrum. These serve as a good example to
explain what gets displayed.
First of all, a GAIA window opens up showing your flat field
frame. If you move the mouse cursor over the image, you can read off
the data value at the current mouse location in the box labeled
Value towards the top right of the numeric display area.
The window should look somewhat like this:

Next a KAPPA window opens. The top left quadrant shows a
histogram of the data values in the flat field frame, the bottom left
quadrant shows you a histogram of data values in the normalised flat
fied, and the bottom right shows you an image of the normalised flat
field. It should look somewhat like this:

You should use the GAIA window and the Histrogram display to
check that your flat field frame has sufficient counts - should be
around 3000 in the brighter areas, and not more than around 4500,
where the array responce starts to go non-linear.
If the count rate is not suitable, then go back to the OT, and edit
the flat field component - adjust the exposure time, or the black body
lamp apperture if necessary.
Next, your Arc spectrum should arrive. Again, the frame will be
displayed in the GAIA window and a histogram of data values
will be plotted in the top left of the KAPPA window. Again, you
should check for a suitable number of counts in the data and adjust
the exposure time if necessary. You should also check for a suitable
number of lines to wavelength calibrate the spectrum. You could try
one of the other arc lamps if you are low on lines. Do these by
editing your program in the OT.
The lower right panel in the KAPPA window shows you the arc
spectrum with an estimated wavelength scale on it (this is derived
from the grating parameters (angle, order etc) reported by the
instrument). Use this to identify a few lines and check that your
wavelength region has been selected with suitable accuracy.
What actually gets displayed
OK, time for a word on what's actually getting displayed.
First up, the GAIA window. We do simply display "raw" data
here, but this isn't necessarily easy to interpret - there may be
multiple integrations in your observation ( see the note on exposures, integrations,
observations and co-adds), and/or the data might be oversampled by
array stepping. The pipeline replaces the display of raw data in the
gaia window with the data in a form more easy to interpret as soon as
it has been reduced into such. In fact, this goes through several
iterations. Usually, you end up with a _wce frame in the gaia window,
which is the most reduced form that single frames go to.
Next up, the KAPPA window. the top left quadrant of this
allways shows you a histogram of data values. The actual image that
this comes from is the same one as what gets displayed in the
GAIA window. Note that especially with the echelle, the
histrogram can be changed significantly by the flat-fielding step, as
echelle flat fields can have significant CVF gradients across them,
and thus when normalised, still contain a range of values. Thus, if
you want to check whether you're saturating, make sure you check a
pre-flat-fielding frame when using the echelle.
The top right of the KAPPA window, shows you the "bgl"
file. This is a frame that shows you how background limited your
observations were. The colour scale for this image goes from 0 to 2,
where 1 is when the Poisson noise from the number of detected
photoelectrons equals the readnoise of the array. For maximum
sensitivity to faint objects, your exposures should be long enough to
make almost all the pixels background limited - ie this frame should
be well filled with values greater that 1. Of course, when observing
bright targets, this is not possible as to do so whoudl saturate the
array on the target.
When you're observing a Flat, the lower half of this window shows
you the histogram of pixel values from the normalised flat field frame
on the left, and an image of the normalised flat field frame on the
right. You shouldn't be able to see large amounts of structure in this
image other than the bad pixel mask which will be apparent.
Other than when taking flats, the lower right panel of the
KAPPA window shows an extraction of rows 139-141 of the lastest
image, with an estimated wavelength scale applied. Thus, this is
useful for checking wavelength regions with arc observation. During
normal observations of a target on the sky, this will show either a
simple extracted spectrum of your target, with no sky subtraction, or
a sky spectrum, depending on whether you're observing in the main or
offset position at that time. This assumes that you are using the
conventional peakup row for your targets (row 140).
The lower right of this KAPPA shows you the y-profile of
your sky subtracted group image, once you have observed sufficient
data to form the group image. Generally, this is useful to see how
well you're detecting your object and to check that you're getting
equal flux in all beams if you're nodding (or chopping) along the
slit. If you see here that your beams are unequal, you may need to
stop and do a 2-row peakup.
Continuing Observations
Once you have completed a pair of main and offset beam
observations, the pipeline forms a group file, into which it subtracts
the main and offset images. This image is then displayed in a seperate
GAIA window, the spectrum extracted from both rows in it is
displayed in the top half of a second KAPPA window and the
y-profile is displayed on the lower left of the first KAPPA
window and mentioned above. Examples:



If this object has been flagged as a standard star (in the OT), then
the pipeline will consult the BS catalogue or SIMBAD to determine the
spectral type (and hence temperature) and V magnitude of the star
(which are displayed in the extracted spectrum title). It will use
these to create a black body model of the star (nb. no interpolation
over stellar features currently takes place), which will be used to
flux calibrate subsequent data.
As soon as you have observed a non standard star tagret in the main
and offset beam positions, a group file will be created and displayed
and spectra will be extracted and displayed as before. An additional
KAPPA window is created to show you the extracted spectrum
(_sp) and flux calibrated (fc) spectra. These are obviously all
updated as more data is observed and signal to noise ratio is
accumulated. Examples follow:

/
The nightlog is a software generated text file, summarising the
observations over the night. This file is created when the pipeline
sees the first data of the night, and a one line entry for each frame
is added as it is reduced by the pipeline. The file is in your
reduced ($ORAC_DATA_OUT) directory, and has the name UTDATE.nightlog,
where UTDATE is obviosuly the current UT date (eg
20000915.nightlog). If you stop and re-start the pipeline, and any
frames are processed twice, they get two entries in the night log
file.
During the day, after you finish observing, a nightlog file is
automatically created in your raw ($ORAC_DATA_IN) directory,
containing the whole nights observing. If the file in your reduced
area gets into a mess, you can create the one in your raw area showing
the observations so far. In a new xterm window, issue the oracdr_cgs4
command to setup the pipeline, then use something like: oracdr
-log sf -skip -resume -from 1 NIGHT_LOG to create the
nightlog file. A full version containing the whole night will be
created here during the day anyway, overwriting any part-night one you
leave there.
Nightlog files are best viewed by stretching an Xterm window to be at
least 132 columns wide, then simply using more to scroll through the
file.
Changing your mind about the recipe
Or - the DR keeps complaining about the flat and exiting
Occasionally, you will find that for some reason, the recipe that you
specified whilst you were using the OT is nolonger appropriate, but
you've allready started taking data with that recipe. This will
usually be brought to your attention by the fact that oracdr will
complain that it's unable to reduce the data and will exit. Normally,
this is due to the absence of a pre-requisite for the recipe that you
have specified. Most of the recipes use a flat field, and will fail if
there is not one availiable. The target recipes that attempt to divide
by a standard star observation and flux calibrate the data will fail
if there is no suitable standard star.
There seem to be a few common reasons why this occurs:
- You're convinced that you have taken suitable
calibration data, but oracdr refuses to use it for some reason.
- Your obsject is about to set, so you intentionally deffered
taking flat, arc and standard star observations untill after your
object frames.
Starting with the first case. Oracdr will tell you why it
is refusing to calibrate, even though it can be a bit crypic
sometimes.
First a note on notation. We refer to flats, arcs and other similar
frames as calibration frames.
The first thing to check is that oracdr knows about the calibration
data which you consider to be suitable. The pipeline keeps an index of
calibration frames, and looks to this index to find a suitable one
when it needs one (eg to flat field some new data). It adds the
details of calibration frames to this index as they pass through the
pipeline, therefor if the frame (say the flat field) hasn't been
reduced by the pipeline - say for some other reason you didn't have
the pipeline running when you took the flat field data, the oracdr
doesn't know about the frame and thus won't use it. Simply pour the
calibration data through the pipeline with a command like oracdr
-list NN where NN is the frame number of say the flat field, then
restart the pipeline on the frame where it fell over. If you did a
sequence like Flat, Arc, star, target, then you might want to simply
re-start the pipeline from the start of that sequence - ie oracdr
-loop flag -from NN.
If the pipeline has processed the calibration data, but still refuses
to use it, examine the output of oracdr to find out why it refuses to
use it. Whe oracdr requires a calibration frame, it works backwards
through it's index file of that type of frame (eg flat fields. The
actuall index is a human readable text file - in this case
index.flat in your reduced ($ORAC_DATA_OUT) directory),
checking the headers of the indexed frames against a rule set to see
if they are suitable. It will use the most recent frame that matches
the rules for use. Typically, the rules file specifies that the
optical configuration must be the same for the calibration and data
frames - ie the grating, order, wavelength, etc etc must match. With
CGS4 there is the added complication that if you move the motors to
change from one optical configuration to another, then attmept to move
back to the original, the optics will be at slightly different
positions, simply due to the inability of the motors to position that
accurately. This effect is such that flats and arcs taken before the
change and change-back are NOT suitable for calibrating
the data. Therefor we have this concept of a configuration
index - the CNFINDEX FITS header. This is simply an integer
which gets incremented each time an optical configuration motor is
moved. We demand a CNFINDEX match between data and calibration frames
to be sure of suing appropriate calibration frames. Therefor if you
took flats and arcs then moved the motors (perhaps something failed
and you had to re-datum, or you accidentally ran the wrong sequence),
you need to re-do your flats and arcs.
OK, now onto the second case where you deliberately don't have the
calibration data because you decided to defer those observations
untill after you'd observed your target (most likely because it's
setting on you, or there's some other time-critical constraint). Now,
or course, it's impossible for the DR to do the normall reduction
sequence, but you still want it to display your data s it comes in, as
best it can, so that you can see how you're doing. The best way to do
this is to specify an alternate recipe wih which to reduce the
data. Note. This doesn't change the recipe specified in the headers of
the data files, it simply forces oracdr to use a specified recipe for
the moment. The recipies you are most likely to want in this situation
are the ones ending in _NOSTD (it it's trying to divide by standard
and flux calibrate, yet you don't have the standard star observations
yet) or the ones ending in _NOFLAT if you don't have a flat field
yet.
Later on, when you come to re-process the data after you've
taken the necessary calibration data, all you have to do is reduce the
calibration frames before the target frames, by means of the
-list option to oracdr, then when you process the target
data, it will use the originally specified recipe, with the correct
(albeit taken later) calibration frames.
Data Reduction Recipies for CGS4
In general, if you based your observations on the template library,
the DR-RECIPE component will be all set up for you, and you won't need
to chnage anything.
The ORAC-DR section of the CGS4 manual provides more info on what
the individual recipies do and the observing sequences they are
appropriate for. If you are unsure or if you think you will need a
recipe that is not provided in our standard selection, you should
contact your support scientist or the CGS4 instrument scientist
well in advance of your observing run.
CGS4 data files under ORAC
Raw files in $ORAC_DATA_IN
The CGS4 raw data files are now stored as starlink HDS
containers. Each file is equivalent to 1 observation, and as
such contains a header component and 1 or more (actually NINT)
integrations. Each integration is stored as an NDF component of the
HDS file.
The filenames are thus: cYYYYMMDD_NNNNN.sdf, where YYYYMMDD is the
numeric UT date (year, month and day) and NNNNN is the observation
number, padded with leading zeros when necessary.
Reduced single frame files in $ORAC_DATA_OUT
First, some conventions:
The filename structure is: (PREFIX)(UTDATE)_(FRAME
NUMBER)[_(EXTENSION)].sdf, where (THIS) is allways there and [THIS] is
optional.
(PREFIX) is the letter 'c' if the file contains data from a
single observation. It is 'gc' if the file contains data from a number
of observations - ie a group.
_(EXTENSION) is used by individual primitives (think of a primitive
as a single step within a recipe) for their output files. The pipeline
its self keeps track of passing these files between primitives, though
all the useful ones are left on the disk at the end so you can look at
intermediate data products if you wish.
For example, c20000410_00123_ff.sdf would be data from a single
observation, number 123, that has been flat fielded (the table later
tells you the _ff means flat fielded).
So is that and HDS or an NDF?
Well, often, it depends; the pipeline passes files between primitives
as either HDS or NDF, depending on which is most appropriate. For
example, with 2x2 sampled data, the raw file will be an HDS container
containg the 4 NDF data arrays from the 4 sample positions. If we're
using a 1x1 sampled flat field, the flat field primitive will flat
field each component, and write out an HDS container containg the four
flat fielded data arrays as NDFs. On the otherhand, if the flat field
is taken with the same sampling as the data, the pipeline will
interleave the samples of both the data and the flat field before
carrying out the flatfielding, thus the _ff file would be a single
NDF.
Of course, some primitives will only write out single-NDF files;
for example, the primitive that interleaves and/or coadds the samples
into one larger image writes out _inc files, and these are allways
single component NDFs.
An additional factor is that later on in the processing, the
pipeline may go back to using HDS containers. This happens for example
when we start extracting beams and spectra from the reduced group
image - HDS containers are used, containing information for each beam
- for example opt-extract profiles, and the spectra themselves, whilst
they pass through the primitives that operate on the individual
extracted spectra before these are combined to give the final spectrum
- for example de-rippling.
File Extensions for single observation files
| Extension |
HDS or NDF |
Description |
| _mraw |
either |
A Modifiable copy of the raw data. |
| _bp |
either |
Bad Pixel Mask has been applied |
| _rnv |
either |
Read Noise Variance added |
| _sbf |
either |
Subtracted Bias Frame |
| _acb |
either |
Added Chop Beams (used for Flats) |
| _scb |
either |
Subtracted Chop Beams |
| _pov |
either |
Includes Poisson Variance |
| _bgl |
either |
How BackGround Limited the integration is |
| _ipm |
either |
Interleave Prepared and Masked |
| _inc |
NDF |
Interleaved and Coadded |
| _ff |
either |
Data has had flat field applied |
| _nf |
either |
Data is a normallised flat field |
| _wce |
NDF |
Wavelength Calibrated by estimation. This is the equivalent of the old ro* file |
| _ss |
NDF |
Sky Subtracted |
This figure illustrates the data flow between primitives for the
reduction of a typical single frame.

Reduced group files in $ORAC_DATA_OUT
The pipeline adds the individual frames into the group as they are
processed. The group number is usually the frame number of the first
frame in the group. Group files are allways single NDFs.
File Extensions for group files
| Extension |
HDS or NDF |
Description |
| No extension |
NDF |
This is simply the difference between all the main and offset beam
images. These are the equivalents of the old rg* files. |
| _oep |
HDS |
The opt-extract profiles |
| _oer |
HDS |
The opt-extract profiling residuals |
| _oes |
HDS |
The opt-extracted spectra |
| _rif |
HDS |
The de-rippling flat fields |
| _dri |
HDS |
The de-rippled spectra |
| _ccs |
HDS |
Cross-Correlated and Shifted. All the beams are spectrally aligned to beam 1 |
| _ccf |
HDS |
The Cross-Correlation Functions from forming the _ccs frames |
| _sp |
NDF |
Extracted Spectrum - the coaddition of all the beams |
| _aws |
NDF |
Aligned with Standard. Spectum is specrally aligned with the standard star |
| _scf |
NDF |
Standard cross-correlation function. The CCF from forming the _aws frames. |
| _dbs |
NDF |
Data has been divided by a standard star (including the standard star black body model). |
| _fc |
NDF |
Flux calibrated. |
Introduction
This guide is not meant to be a complete cgs4 data reduction
manual. It outlines the steps a typical UKIRT/CGS4 observer will want
to carry out on their data after the observing run. These steps could
be carried out using many different data reduction packages, for
example starlink or IRAF.
I will assume that you have a nightlog file from the orac-dr
pipeline, and that you have the _wce files for each science frame,
which will consitute the starting point for this guide. I will also
assume that the _wce files you have have been processed satisfactoraly
by the pipeline, and have been flat-fielded. This will usually be the
case.
Stage 1: co-add into groups.
With reference to your nightlog, select the _wce files that
consitute each "group" of observations. A group would typically be an
observation of one target at one instrument configuration.
Inspect each of these frames and check for any obvious defects. If
you discard and frames, be sure to discard the same number of
main-beam and offset-beam frames. The main-beam frames will have 0,0
as the RA and DEC offsets in the nightlog, the offset-beam frames will
typically be 11.74 arcsec away from 0,0. If you were nodding to blank
sky, the offset will be much larger, and the offset-beam frames will
also be SKY frames.
Form a group frame which consists of the sum of all the main beam
observations minus the sum of all the offset beam observations.
Do this for each group of observations you are going to reduce at
this time. You should at least have a standard star group and a target
group at this point.
In these group files, the "sky" areas should be relatively free of
sky-lines. Sky line residuals up to a few tens of counts can be
removed (see later). You should have a positive band acorss the image
representing the positive image of the spectrum of the target. If you
were nodding along the slit, you should also have a negative band
across the image representing the offset-beam negative image of the
target spectrum.
If your observations spanned a long time interval, ie a large
change in position of the telescope as you tracked the object across
the sky, then you should consider adding your observations into
"sub-groups" of an hour or so's data, extracting the spectra from each
of these seperately, then doing a cross-correlation to determine any
shift due to instrument flexure, and applying this shift before adding
the spectra together
If you wish, at this point, you can subtract off the residual sky
lines. Use your favourite DR package. You should mask off the areas of
the image containing the target spectra, then have the package fit a
low order polynomial to the residual sky flux in each column. The
polynomial fit should then be subtracted from the whole image,
including the areas where the target spectra are, obviosuly.
Stage 2: Extracting the spectra
Extract the +ve and -ve images of the spectrum seperately. You can
use optimal extraction if you like, and if you used the same nod
distance for the standard star, you can use the standard star
observations to generate the profiles with which to optimally extract
the target spectra. After you extract the -ve spectrum, multiply it by
-1 to make is positive. Keep the profiles that you use around for
later use.
Check for a shift between the specta. This needs to be done on
something bright; I suggest doing it only on the standard star
spectra, however if your target is brighter than your standard (you
might have been doing monitoring, or looking for weak lines on a
strong continuum etc), then you could do it to the object spectra too.
Basically, cross-correlate the main-beam and offset-beam
spectra. If you get a shift of more that say 0.2 pixels, shift the
offset beam spectrum to match the main-beam spectrum. If you're only
measuring this from the standard star spectra, apply the same shift to
the offset spectrum from the target data too.
Add the main and offset beam spectra together. This forms the
"observed spectrum" of each object (be it the standard star or your
target).
It'll probably keep things simple if you normalise each spectrum by
its exposure time at this point.
Stage 3: Wavelength Calibration
You'll have noticed that the data allready have a wavelength
scale. This is approximate - it's from an estimation based on the cgs4
motor positions. You should now do an accurate wavelength calibration
using the arc spectra you took.
Look at the nightlog file and find the _wce frame for the arc-lamp
observation you're going to use. This will normally have been taken
shortly before the 1st standard star at that confiuration. If several
are availiable, use the one at closest telescope position to where your
target was observed, to minimise effects from instrument flexure.
Extract a spectrum from the arc frame, using the same rows you used
for the target frame main-beam. If you did an optimal extraction, use
the same profile as you did for the main-beam.
Use the arc-lamp maps on the web to identify the lines in your arc
spectrum. Use the wavelength calibration routine in your data
reduction software. The principle of these routines is that you tell
the software the exact wavelength of a number of lines. The software
measures the position of the peaks of these lines and comes up with a
function to relate the wavelength to pixel number. You should use as
low-order function as gives a reasonable fit. A 3rd order polymonial
is usually sifficient, though 5th order is sometimes usefull,
especially if you have a large number of identified lines in your arc
spectrum.
Apply the calibration you've just generated to the target and
standard star spectra you extracted earlier.
Stage 4: Flux calibration
There are many ways to do this. This what I do for a reasonably
simple case
Create a black-body model of the standard star
Look up the spectral type of your standard star. Each spectal type
corresponds to a black body temperature. There's a table of these on
the ukirt web pages. Have your DR software generate a black body
spectrum at the appropriate temperature, on the same pixel and
wavelength scale that your standard star is on.
Next, deduce the flux of your standard star at some wavelength
which is on your spectrum. Usually you'd use the band centre, and
you'd deduce the flux from either the known magnitude of the standard
in that band, or the magnitude in some other band, determining the
colour from the spectral type. Scale your black body spectrum so that
it has the correct flux at this wavelength.
Create a sensitivity spectrum
Take the black body spectrum you've just made (which you've scaled
to be in flux units, say W/m2/um), and divide it by the "observed
spectrum" of your standard star (which is probably in counts per
second or similar). This gives a sensitivity spectrum - ie the value
at each wavelength is the flux needed to produce 1 count per second at
that wavelength.
BUT, there's a small problem in that the standard star probably
doesn't really have an exact black-body spectrum. It more than likely
has some lines in it. You should be able to identify these lines in
the sensitivity spectrum. Ask a local expert or your support scientist
if you're not sure. Different types of star have different lines. try
starting off looking for hydrogen recombination lines.
When you find a feature in the sensitivity spectrum that you think
arrises from a line in the star, interpolate over it using the
"continuum" either side of it.
Apply the calibration
Simply multiply your sensitivity spectrum by your "observed
spectrum" of your target, to get a flux calibrated spectrum of your
target.
|