IRCAM3 manual
System Gain, Noise and Linearity
System Gain
The IRCAM/TUFTI system has a gain of 6 electrons per ADU.
System Noise
Typically STARE mode observing results in a noise of 56 electrons
and NDSTARE 47 electrons. Hence to be background limited, counts on
background need to be 2-3 times greater than the square of the readout noise,
or, for NDSTARE, counts need to be around 700 ADU.
Linearity
Significant non-linearity sets in for IRCAM
at levels below those stated in the old manual. For Standard readout bias
settings, data will be 2% non-linear at 8000 counts and 5% at
12000. To retain linearity keep your counts on any target of interest below
8000 counts with Standard
ND-Stare readout (or 33000 in Stare mode).
The linearity correction for NDSTARE images is given by
true/measured = 1 + 3.3e-06(measured)
A
postscript plot is available
showing the linearity correction for NDSTARE IRCAM images (add 25,000
to the counts for the corresponding STARE level). ORACDR will
correct standard readout data for non-linearity.
In Deepwell mode, used in the thermal, the array also
behaves non-linearly, however in this regime counts are
dominated by the thermal background and your calibrator and target will have
very similar total counts
in the raw frames, hence the problem is much reduced. Use of Deepwell also
removes the sharp saturation effect seen at 16,000 counts in standard NDSTARE
images; hard saturation instead sets in around 30,000 counts. The linearity
correction for Deepwell images is given by
true/measured = 0.8298 + 8.186e-05(measured) - 1.224e-08(measured)^2 +
9.224e-13(measured)^3 - -3.207e-17(measured)^4 + 4.164e-22(measured)^5
A
postscript plot is available
showing the linearity correction for Deepwell images (add 25,000
to the counts for the corresponding STARE level). ORACDR does
not correct Deepwell images for linearity as these frames are dominated by a
high and nearly constant background level.
|