L-band polarimetry is now available with TUFTI. Following the
recent upgrades to IRCAM (addition of a cold snout and re-scaling of
the pixels to 0.08"), IRCAM is now optimised for use in the thermal IR
(L and M-bands) and has acquired a new name, "TUFTI". Details of
polarimetry options with TUFTI are given in the pages listed below.
Users of IRPOL2 with IRCAM/TUFTI may also find the IRCAM3 Polarimetry
manual written by Ant Chrysostomou of some use. It contains a number
of very helpful tips and comments concerning imaging polarimetry in general.
A postscript version is available
here.
Polarimetry with IRCAM/TUFTI: Data Acquisition
Getting started - IRPOL setup
IRPOL2 comprises a half-wave retarder (the waveplate), a focal-plane
mask and (internal to IRCAM/TUFTI) a Wollaston prism (the focal plane
mask and half-wave retarder will be installed by the Telescope System
Specialist, or T.S.S., at the beginning of the night). Separate
waveplates are available for JHK, L and M-band polarimetry. The prism
splits the incoming radiation into orthogonally-polarised e- and
o-beams. With extended sources (greater than about 5 arcsec in
declination), these beams would overlap on the array. Hence the need
for the mask. At present a warm, focal-plane mask must be positioned
in front of the IRCAM/TUFTI "snout". For thermal imaging, the
SILVERED masked should be carefully positioned at the spot marked on
the bench. IMPORTANT: Take a few exposures
of blank sky or even the inside of the dome at the beginning of the
night to check that the mask is properly positioned; a raw image with
the mask properly positioned should look something like that shown in
Fig.1 below. Any hairs or fibres on the mask will appear "resolved" in
both e- and o-beam images (Fig.1. shows "nice" examples of this);
these can be blown off the mask with a can of pressurised air.
Fig.1. A typical raw L-band IRCAM+Pol image with the mask well
aligned.
An IRPOL image is split into 4 east-west strips. The two northern
strips represent "extra-ordinary" and "ordinary" images of the same
field viewed through the north slot in the mask; the two southern-most
strips are o- and e-beam images of an adjacent field viewed through
the south slot in the mask. Note, however, that on all image
displays the IRCAM field is rotated CLOCKWISE through
89.0o. Consequently, the east-west o- and e-beam images are
rotated, as in the example above.
The reduction recipes assume that the star is located in
the NORTHERN (or right hand) half of the array. The DR recipes extract
object images from this area only; the southern e- and o-beam images
are used to calculate sky values. The usable area of each E-W beam on
the array (i.e. the size of each horizontal strip) measures
approximately 20 x 4 arcsec.
Observing programmes - the ORAC-OT and ORAC-OM
IRCAM/TUFTI imaging-polarimetry data are acquired in much the same way
that UFTI-pol data are obtained. A description of how to use the ORAC
Observing Tool (OT) to prepare observing sequences, and how to
observer via the ORAC Observing Manager (OM) and Sequencer are
available in the UFTI +
IRPOL Data Acquisition pages.
When preparing L-band sequences for IRCAM+IRPOL, please bear in mind
that very short integration times will be needed, typically less that
1 second. The IRCAM field of view is also much smaller than that of
UFTI. Recommended Template Sequences, specific to
IRCAM/TUFTI, are available in the Template Library of the ORAC-OT.
Point Sources and Extended Objects
It is recommended that observers repeat observations of both point
and extended sources several times. This will give the observer an
indication, via internal statistics, of the accuracy of the final
result. With IRCAM/TUFTI, a source that fits comfortably within the
limited field of view of the "e-beam" and "o-beam" images on the array
(see Fig.1 above) may be regarded as a point source. If the user wishes to
retain spatial information and avoid bad pixels, then a small 3 point
jitter pattern mosaic is recommended, with say 5 arcsec east-west
offsets between each mosaic tile. The "Pol_Jitter then Angle
(IRCAM)", and "Pol_Angle then Jitter (IRCAM)" sequences in the
Template Library are provided for such an observation.
For extended sources which are more than ~ 5 arcsec in declination, a
"Pol_Extended" sequence is provided in the Template Library that
allows the user to nod between target and blank sky (at each of the 4
waveplate positions).
Flatfields
With near-infrared astronomy, a good flatfield is of course always
a high priority, although for polarimetry the flatfield is in
principle independent of the final result, since it is the
differences between observations which are measured. However,
one can only safely assume this if the same pixels are compared each
time. Image motion, for whatever reason (some degree of dithering is
recommended to account for bad pixels) will mean that the images must
be shifted to spatially register them, rendering the assumption of
flatfield independency invalid. Once again, a good flatfield is - as
in most cases in astronomy at UKIRT - crucial!
There are a variety of methods which the observer can employ to
obtain a flat field. At present we provide two Template Sequences that
can be used to obtain flat field frames; the first combines images
taken at different waveplate angles to make a "master-flat", while the
other - which is more time consuming, though is possibly more accurate
- computes flats at each of the waveplate angles separately. In
either case, with IRCAM/TUFTI at L-band, getting sufficient
signal-to-noise on the array will not be a problem!
Note also that, because a focal plane mask is used with TUFTI, the
corresponding flatfield frames must also be obtained with the focal
plane mask in place. This is because the flatfield response of the
array is sensitive to the state of polarisation of the radiation, and
since the prism and mask always ensure that only e- or o-beam
radiation is transmitted onto a given area of the array, then the
flatfield response of that same area of the array must likewise only
be measured in either e- or o-beam radiation. Once again, the purpose
of the mask is to ensure that the two beams are kept separate on the
array.
Changing Waveplates During the Night
For L-band polarimetry a zero-order waveplate is available. Note
that, if near-IR polarimetric imaging is also required for your
project, there are a limited number of near-IR filters installed in
IRCAM/TUFTI at any one time. Consequently, JHK polarimetry can be
obtained with IRCAM/TUFTI, provided the filters are not in the same
filter wheel as the prism. This might be a more efficient option than
switching between IRCAM/TUFTI and UFTI, although note the much larger
field of view associated with the latter.
The L-band waveplate is warm (it sits in an extendable arm within the
instrument support unit - ISU2). We also have an M-band waveplate,
though tests suggest that it is unusable with IRCAM/TUFTI because of the high
background. The waveplates are large (95mm diameter) to accommodate
the full IRCAM/TUFTI field of view. This makes them VERY EXPENSIVE!
Therefore, only certain UKIRT staff are allowed to handle the
plates. If your observing programme calls for a change of waveplates
during the night, then you should inform your support scientist of
this. All the TSSs know the procedure and
safety aspects of changing waveplates, and they should be the ones to
do this during the night if required. Observers should NOT attempt to
change waveplates!
The polarisation efficiency of IRPOL at 3-4 microns has been measured with
CGS4, and found to be wavelength dependent though in excess of 90%.
Users should consult
these pages for further details.
To calibrate measurements of polarisation position angle with
IRCAM + IRPOL, deep observations should be obtained of a polarised
standard. As a guideline, note that from recent observations of the
polarised standard HD38563C with UFTI we measured the degree of
polarisation and position angle at K to be 2.10(+/-0.09)% and
87(+/-6)o. These compare with catalogue values of
2.21(+/-0.55)% and 78(+/-17)o (Whittet et al. 1992, ApJ,
386, 562). The data suggest that a correction of -9o
should be applied to p.a. measurements.
Measurements with IRCAM3 -- prior to it conversion to IRCAM/TUFTI in
summer 1999 -- indicated the need for a similar correction of
-6o.