|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
In theory, there is only one DM screen you need to run in order to observe
successfully with Michelle. However, we will describe a few others in case
they are needed for troubleshooting. At the very minimum, you will need
to run the mich_oper.dl screen which is launched by hitting the "mich_oper
screen" button in the above menu. The following will launch:
Enabling the array: At the beginning of the night, or when Michelle
is about to be used for the first time, you must enable the array using
the "Enable" button. The array control state will turn green from yellow,
and the channel A & B temperatures will rise. Remember to disable
the array at the end of the night or when you have finished using Michelle. Closing the shutter: The Michelle shutter, situated in the warm
calibration unit, is essential for the safety of the KBr window since it's
hygroscopic. Although the shutter will open automatically when you run
the first Michelle sequence, you must remember to shut the window
when you have finished using Michelle, if the outside humidity reaches
80%, or at the end of the night. To do this, simply click on the
"flat/cover"
button and then hit "START". Although the above describes the bare minimum needed to run Michelle,
it is recommended that you also run the STATUS screen (see section 2) and that you also run the WFG and CCS serial
consoles (see section 3). Other useful DM screensCCS status: The mich_oper DM screen can be used as a starting point to launch a variety of useful screens. Probably the most useful in order to monitor the status and Health of the Michelle mechanisms is the CCS Status display available by clicking on the "status" field in the mich_oper DM screen.
The cryostat control system: The CCS display allows you to access individual DM displays for each mechanism, enter settings and move them. The display also allows you to quickly force the mechanisms to find datum but only after a reboot of the CCS. It is launched by holding down the right hand mouse button on the "other displays" field and selecting "CCS".
After a reboot of the CCS, you can force the mechanisms to find datum
by simply hitting the "apply" button. Each mechanism comes back with pre-determined
settings after a reboot and this button simply sends a request to each
mechanism to move to those settings. Since the mechanisms lose their reference
position after a reboot, they are automatically forced to re-find their
datum positions before moving to the desired setting. In normal operations,
the mechanism will probably not move if you hit this button since each
mechanism will probably already be at the correct setting. NB. If the CCS has been rebooted, sending and executing a Michelle
sequence in the OCS software will also result in the mechanism finding
datum. If, for some reason, you need to move an individual mechanism, say, for
troubleshooting purposes, hold down the right hand mouse button over the
relevant button and select the option you want. In most cases, this will
be the first menu item, but buttons such as the detector and imaging buttons
also give you a choice of two separate motors, For the detector these are
the detector focus and translation, and for imaging this is the injector
and extractor. As an example of what can be done, we'll look at the grating
DM screen.
NB. Before moving any mechanism using the individual
mechanism DM screens, you should contact Tom Kerr for assistance. The top third of the screen simply displays the current settings,
the bottom two thirds is where the action takes place. Most buttons and
fields are probably obvious. The set wavelength buttons allow you to specify
whether you are entering a wavelength or a mechanism step position. The
datum search buttons will either force the mechanism to find datum ("force")
or allow the software to use its intelligence and decide whether the mechanism
needs to find datum or not ("discretionary"). In just about every case, discretionary
should be selected. The order selection buttons specify whether the software
determines the grating order (automatic) or whether the user will specify
it. The grating can be selected using the 5 buttons on the left,
with the wavelength (or step position) being entered in the brown field
in the middle of the display and the order (if necessary) being specified
on the right. Note that zero order cannot be selected and that a step position
has to be entered manually for this. For the LowN grating, zero order is
found at 4572 steps. Power is automatically applied to the motors when the START
button is clicked and the wl_power and drum power buttons will toggle.
Note that you must hit the start button in order to move the gratings to
the desired setting, they will simply sit at their original position if
you do not. Emergency stop: There have been occasions with other
instruments that an observations is accidentally started with so many coadds
that the integration will take several minutes, or even hours, to stop.
Hitting the stop asap button in the sequence console doesn't help, as the
exposure will only stop after the set number of coadds have been done. The
same is true for Michelle, although it is very unlikely such an observation
will be created in the first place. In case you do need to abort an observation in an emergency,
you can do this via the Edict DM screen available by holding down the right
hand mouse button on the "other displays" field in the mich_oper screen,
and selecting "Edict".
Simply hit the "abort" button to stop an integration (the button is not clear in Fig. 6, it's the dark coloured button immediately underneath the observe button). NB. The effect that aborting an integration will have on the OCS software is not tested - it's likely it will cause the OCS to go into an error state and that you'll have to restart the OCS software. The Edict and CCS serial consolesBoth Edict (the array control software) and the CCS (cryostat control system) have serial consoles which can be used to enter commands, monitor the software and be used to reboot the systems. Although not absolutely necessary, it is recommended that the WFG and CCS serial consoles are launched as a matter of routine since they are useful to have in case there are any instrument problems. Both of these consoles can be launched from the michelleMenu screen by selecting the "CCS serial console" and "WFG serial console (only)" buttons.
In normal operations, both consoles will display logging
information and software settings relevant to both the CCS and Edict systems,
therefore if there is a problem of which the cause is not clear, you should
glance at these consoles to see if an error message is displayed. Occasionally,
you may also need to open all the edict serial consoles
while troubleshooting to see if any errors are being reported by the slaves.
These can be launched using the "All Edict serial consoles" button on the
michelleMenu display. If systems need to be rebooted via software, you can do this
via the serial consoles. Simply type "reboot" in either the WFG or CCS
serial console. Remember - rebooting in the WFG serial console will reboot
Edict only, and rebooting in the CCS serial console will reboot the CCS
only. Please be sure to close all the serial consoles when you
are finished with Michelle. Target acquisition and peakupOne big difference with Michelle in 04A compared to the last time it was at UKIRT is the ability to acquire a spectroscopy target in imaging mode without the need for a painful peakup. Although this hasn't been fully tested yet, initial tests at night have been successful.Instructions:
Run a normal spectroscopy sequence until you reach the target acquisition
command. The sequence will automatically pause at this point. When paused,
click on the MOVIE button and the movie dialogue window will be displayed.
Simply click on the START button and after a few seconds a chopped image
will be displayed of your target area in the quick look tool. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||||
|
|
Contact: Tom Kerr. Updated: Wed Oct 6 13:29:24 HST 2004 Return to top ^ |