Plots are available showing timescales for
warming up and
cooling down UFTI. For a normal cycle UFTI requires
30 hours to warm up and 40 hours to cool down. Due to
differential contraction the shutter and filter wheels
should not be moved until they are below 100K. Cold vacuum
pressure should be around 2e-6, and initial pressure
on a cooldown is usually 1e-3.
Here as a copy of the
temperatures epics display,
showing nominal temperatures for UFTI and indeed all of the instruments
(current as of Nov. 2001; these stable values
slightly higher after change of
cold-head in Spring 2001).
|