UIST logo

Instrument Apertures and Foci

UIST logo


ENG History:

NOTE: Details on how to measure the UIST apertures and Focii are given on the main QA-night web page.


Apertures (in arcseconds)
(Last update of this table - 20 June 2005)
All apertures to be measured on QA night, please
 
Mode to 20040622to 20040820to 20050628 current
Imaging 0.12 -5.77,-1.07 -6.37,+2.39 -6.83,-3.29 -7.43, -1.59
Imaging 0.06 -2.59,+0.75 -3.31,+4.31 -3.56,-1.24 -4.28, 0.02
Spectroscopy +6.65,-0.75 +6.73,+5.03 +6.37,-0.69 5.77, 1.11
IFU +0.69,-1.07 +0.83,+4.67 +0.67,-1.29 0.07, 0.41
Polarimetry +24.43,+1.08+24.69,+6.21+24.37,+0.41 23.77, 2.11

Notes:
1. Use the same aperture for imaging-polarimetry and spectro-polarimetry.
2. The arcsec values are simply the mm values x -1.524
3. The first value is E-W, the second N-S, so in the rotated UIST frame the first value is along the slit (up and down, or in Y in pixels - see below)
4.The difference between apertures should be fixed (so strictly speaking we only need to measure two apertures - imaging at the two camera plate scales). Approximate relative offsets are:

 - Spectroscopy = Imaging(0.12arcsec) + 13.2, Imaging(0.12arcsec) + 2.6 
 - IFU          = Imaging(0.12arcsec) +  7.5, Imaging(0.12arcsec) + 2.0
 - Polarimetry  = Imaging(0.12arcsec) + 31.2, Imaging(0.12arcsec) + 3.7 

 

Location on array
Use these numbers for imaging acquisition AND aperture measurements

(Last updated/checked - 30 June '05)
Apertures highlighted in red to be measured on QA night, please
 
Mode Location on array, (x,y)
Spectroscopy, 1 pix slit 506.4 590
Spectroscopy, 2 pix slit 505.2 590
Spectroscopy, 4 pix slit 502.2 590
Spectroscopy, 5 pix slit 501.5 590
Spectroscopy, 7 pix slit 507.7 590
X-disp Spectros, 3 pix slit 503.5 541
X-disp Spectros, 5 pix slit 503.4 541
X-disp Spectros, 7 pix slit 504.0 541
Spec-Pol, 2 pix slit 510.6 740*
Spect-Pol, 5 pix slit 511.6 740*
IFU 499 540
Imaging, 1024x1024 (0.12") 480 480
Imaging, 1024x1024 (0.06") 480 480
Imaging, 512x512 224 224
Imaging Polarimetry 511 740 *

Notes:
* The aperture for imaging-pol and spec-pol is the same; this should be measured WITHOUT the prism in the beam. This aperture puts the star; (1) in the centre of the top aperture in the imaging-pol-mask, or (2) in the centre of the top slit in the 5-pixel spec-pol-mask. The E-beam image/spectrum of the star will then be on pixel 510,823; the o-beam data will be on 510,655 [for H-band]).

 

Additional offsets during source acquisition
(Last update of this table - 8 July '05 - for new IJ/JH)
 
Grism Filter used Additional offset
by TSS at telescope
IJ J-spec right 0.50 arcsec
JH J-spec right 0.50 arcsec
Short_K B2 left 0.05 arcsec
Long_K B2 left 0.90 arcsec
KL B2 left 0.90 arcsec
Short_L 3.4cont left 0.10 arcsec
Long_L BrA right 0.17 arcsec
M BrA left 0.15 arcsec

PLEASE NOTE: UPICK automatically applies the above offsets. TSSs do not have to adjust the source position after imaging acquisition in the thermal (this is why upick puts the star on a slightly different pixel than is listed earlier on this page).


UIST Focii

The plot below shows fine-focus measurements taken on QA nights. Note that not all focii are measured on a given day (so some values are missing). The dates marked "UIST off scope" indicate when the instrument was physically taken off the telescope for a period of time. Thus, after one of these periods values might be expected to change...

  UIST focii

Pink - 0.12 imaging focii; Blue - Spectroscopy focii; Yellow - 0.06JHK focii; Light Blue - 0.06LM focii; Purple - IFU focii; Brown - Pol focii.
NOTE: 0.12" imaging and spectroscopy focii should be the same.

Values being used at the telescope (i.e. NOT necessarily the last value measured) are given in the table below

 

Telescope FF Currently In Use
(Last update of this table - 29 June '05)
 
Mode Focus, mm
Imaging, 0.12" (& Spec) 0.83
Imaging, 0.06"JHK 1.02
Imaging, 0.06"LM 0.70
IFU 1.0
Pol 0.78
Notes:
The 0.12" imaging and spectroscopy foci should be about the same.
The 0.06"JHK focus should be MORE than the 0.12" focus by about 0.19mm.
The 0.06"LM focus should be LESS than the 0.12" focus by about 0.03mm.
The IFU focus should be MORE than the 0.12" focus by about 0.21mm
The Pol focus should be MORE than the 0.12" focus by about 0.03mm


Earlier apertures and focii are available on previous versions of this web page. The date corresponds to when an update was made:
20050628
20040622 | 20040522 | 20040402
20031220 | 20031124 | 20031003 | 20030903 | 20030731 | 20030424 | 20030403 | 20030131 | 20030106


Previous: Grism Wheels | Up: UIST Engineering | Next: UIST Engineering