UIST Spectroscopy
|
Gain and Linearity
|
|
Much of this is NOT required reading for observers; a linearity correction is currently
being applied in the ORAC-DR pipeline.
The information below pertains to the new ARC (formerly SDSU)
controller commissioned with UIST in December 2006. For numbers
specific to the old Edict system, please contact the instrument
scientist.
Bias Voltages and Saturation
The detector reverse bias is automatically selected by the low-level
software. With NDSTARE readout (with full and sub-array imaging and
all spectroscopy modes) 600 mV is used; for thermal imaging (THERMAL
ND and THERMAL CDS) 900 mV is used. The latter gives increased well
depth and better linearity, which are important with the high
background flux encountered at L and M.
The array goes into hard saturation at around 21,000 counts (~135,000
electrons) with NDSTARE; in the thermal saturation occurs at 33,200
counts (~210,000 electrons). See below for details...
Dark Current
At 600 mV reverse bias the dark current is ~0.1 e-/pixel/second.
At 900 mV a value of 0.4 e-/pixel/second has been measured.
Pixel Settling Time and Output Coupling
Performance in the thermal is improved by increasing the reverse
bias, which increase the full well depth, but also by reducing the
readout speed. In thermal readout modes video processing has been
minimised so that the pixel processing time is dominated by the pixel
settling time. 2.5 microseconds is the minimum pixel settling time
(limited by the fibre optics speed). However, as the settling time is
reduced, output coupling
increases. That is, the fraction of signal on an output channel that
remains when the channel is next sampled. This produces a ghost image
8 columns along the detector. The output coupling resulting from a
2.5 microsecond pixel settling time is shown here.
For non-thermal readout a pixel settling time of 7.1 microseconds
has been adopted. This leads to an output coupling factor of only 0.44%.
In the thermal, a settling time of 4.0 microseconds is used, which
gives output coupling of 1.16%.
Gain and Linearity
Tests with the ARC controller suggest that there is no single value
for the system gain. Rather, this seems to vary with flux level, from
around 5.75 electrons per data number (per count) at ~10% full-well, to
~6.65 electrons per DN at ~90% full-well with 600 mV reverse bias
(non-thermal readout). Similar behaviour is seen with 900 mV reverse
bias (the gain is slightly lower because of reduced detector
capacitance).
Linearity curves for the 600 mV (non-thermal) and 900 mV (thermal)
bias settings are shown below. The blue lines give the linearity in
data numbers; the pink lines give linearity after correction by the
flux-dependent system gain level noted above. This correction results
in a linear plot.
In the thermal, it is recommended that the
same exposure time (and/or counts on the source and sky) is used with
the target and the photometric standard star.
|
|
Linearity - 600 mV Non-thermal readout |
Linearity - 900 mV Thermal readout |
|
|
Residuals - 600 mV Non-thermal readout |
residuals - 900 mV Thermal readout |
Click for an expanded plot
NOTE: use the right-hand (900 mV) plots ONLY with thermal imaging modes.
If you're unsure about your data, check the DET_MODE fits header:
900mV is used with modes that end in 'T', e.g. ND1T, CDS1T, etc.
Interpixel Capacitance
Finally, there is very little interpixel
capacitance coupling on the UIST array. Approximately 10% is
transmitted to the surrounding 4 pixels, either side in a row and top
and bottom in a column (with both 600 mV and 900 mV reverse bias).
Data and analysis courtesy of David Atkinson, UKATC (Winter 2007).
|