UIST Spectroscopy
|
Spectroscopy: Slits
|
|
Introduction
In spectroscopic mode UIST will only be used with the 0.12"/pixel
plate scale. The instrument has a 16-slot slit wheel which contains field
stops, polarimetry masks and slits for spectroscopy. The latter are
listed below (see the Engineering pages for the full slit-wheel
populations).
Which slits with which grism?
Please note that, because of focus limitations, new slits have
been installed in UIST for use with ONLY the IJ and JH grisms. In the
OT these are called 2_pix_wide_120s and 4_pix_wide_120s.
The standard slits (1_pix_wide_120, 2_pix_wide_120, 4_pix_wide_120)
should NOT be used with the IJ or JH grisms; likewise the
2_pix_wide_120s and 4_pix_wide_120s should NOT be used with the other
grisms. Users will find that an incorrect combination can not be
selected in the OT.
1 March 2003 - Present
 
| Slit |
Name in UKIRT-OT |
Description |
Use with... |
| Long, 1-pixel-wide slit |
1_pix_wide_120 |
0.12arcsec x 120arcsec |
all grisms except IJ and JH |
| Long, 2-pixel-wide slit |
2_pix_wide_120 |
0.24arcsec x 120arcsec |
all grisms except IJ and JH |
| Long, 2-pixel-wide slit |
2_pix_wide_120s |
0.24arcsec x 120arcsec |
only the IJ and JH grisms |
| Long, 4-pixel-wide slit |
4_pix_wide_120 |
0.48arcsec x 120arcsec |
all grisms except IJ and JH |
| Long, 4-pixel-wide slit |
4_pix_wide_120s |
0.48arcsec x 120arcsec |
only the IJ and JH grisms |
| Long, 5-pixel-wide slit |
5_pix_wide_120 |
0.60arcsec x 120arcsec |
all grisms |
| Long, 7-pixel-wide slit |
7_pix_wide_120 |
0.84arcsec x 120arcsec |
all grisms |
| Pol Mask, 2-pixel-wide slit |
pol_mask_spec_2 |
0.24arcsec x 20arcsec (x 2) |
all grisms except IJ and JH |
| Pol Mask, 5-pixel-wide slit |
pol_mask_spec_5 |
0.60arcsec x 20arcsec (x 2) |
all grisms |
Spectral Resolution
The spectral resolution with each grism is given on the next page. Generally, the 2_pix_wide_120 slit
gives roughly twice the resolution of the 4_pix_wide_120 slit. This
is not the case with the IJ and JH grisms, however, where only
moderate improvement (by ~30%) in spectral resolution is achieved with
the narrower 2_pix_wide_120s slit.
The 1 and 2-pixel slits both give essentially 2-pixel
resolution. See e.g. these portions of HK spectra: 1-pix spec, 2-pix spec.
Spectro-Polarimetry
The two polarimetry masks, pol_mask_spec_2 and pol_mask_spec_5,
each have two aligned, 20"-long slits, separated by about 25". These are used
for spectro-polarimetry of an object and adjacent blank sky (see the Polarimetry
web pages for further details.).
Recommendation on Slit Position Angles:
If a specific slit position angle is not required, we recommend
using an east-west slit (i.e. p.a. set to -90 degrees in
the ukirt-ot). When acquiring the source, the "acquisition image" will
then appear with the same orientation on the Movie display as a normal
"imaging-mode" image (i.e. N to the left and E up). Any tracking
errors will also move the source "up and down" the slit, rather than
"off" the slit (though such errors are extremely unlikely!).
When a specific position angle is needed, for angles greater
than 90 degrees we recommend using NEGATIVE VALUES for the position
angle, i.e. for an angle of 110 degrees E of N, enter -70 degrees in the
OT. This will help the TSS with source acquisition.
With a position angle of -90 degrees, the TOP of the slit will be to
the EAST; 0 degrees puts the top to the south. A table showing the
relationship between position
angle and slit orientation on the sky is available in another section.
|