|
UKIRT Annual Report 1998
THE UNITED KINGDOM INFRARED TELESCOPE
ANNUAL REPORT
1998
4. Approved Programme
4.1. Completion of the UKIRT Upgrades Programme
At the beginning of 1998 the Upgrades Programme was formally complete
except for insulation of the dome floor and the primary mirror cooling
system. During the year the benefits of the upgrades were dramatically
demonstrated by the results of a seeing monitoring programme which
confirmed that UKIRT's images now only rarely approach anything as bad
as 1 arcsec FWHM; users now tend to regard 0.7" as ``poor'' seeing.
4.1.1. Telescope Optical Performance
Between 13 February and 29 September 1998 a standard protocol was used to
estimate the delivered K-band image FWHM in a uniform way, ideally twice
per night, whenever the imager IRCAM3 was in use. A total of 143
measurements were secured, 41 of them in September, when IRCAM was
continuously on the telescope.
About 75% of the images in the resulting data-set were taken at the
so-called x 1 pixel scale (0.28"/pixel) and about 25% with the
x 2 magnifier
(0.14"/pixel). The two image size distributions differed notably on
account of undersampling of the images with the larger pixel scale, and
sets of consecutive measures with the two scales were acquired to
construct a correction curve. This was used to correct each observed image
FWHM to a ``fully-sampled'' equivalent. The resulting corrected distribution
was remarkable in that a significant fraction of the FWHM measurements
appeared to be within 0.05" of that of a diffraction limited image
(nominally
0.108" at 2.2 µm with UKIRT's aperture and central obstruction).
It was not initially believed that UKIRT really was regularly offering
images near the diffraction limit, especially as this conclusion
depended on a correction for sampling effects by a factor of more than
two. However, on 8 September 1999 Antonio Chrysostomou and Thor Wold were
able to secure consecutive well-focussed images with no magnifier
( x 1) and
with the x 2 and x 5 magnifiers. The latter, yielding pixels
only 0.057"
square, could fully sample images quite close to the K-band diffraction
limit.
The results not only confirmed the undersampling correction
curve, but yielded images of extraordinary quality. The mean of the FWHM
of three 20 second exposures was 0.178"; the best image had formal
FWHM =0.171" with a Strehl ratio of about 0.25 (i.e. the peak intensity
in the image is
25% of that in a perfect image sampled with the same pixel distribution).
We believe that these may still be the best (sharpest) images yet secured
by a ground-based telescope without the use of high-order Adaptive Optics
(i.e. higher than tip-tilt).
These images are all the more remarkable because the TSS
considered that the night, while good, was not of the very best. This was
in fact consistent with the histogram of corrected images: the images
with the x 5 magnifier fall in the next to smallest populated bin; at
least 10% of the sample images appear to be better than these.
Over the period of the seeing measurement campaign
the median corrected image FWHM was 0.433". 90% of image
FWHMa were below 0.86" and 10% were less than 0.15", i.e.
within 0.05" of
the diffraction limit. Over the month of September the median corrected
image FWHM was 0.265" and the worst image had FWHM 0.563".
The seeing campaign was terminated when the new imager UFTI had its first
light on 30 September 1999. While the sampled period in 1998 may have been
of above-average seeing overall, UFTI, with its 0.091" pixels, has
since yielded many images with FWHM between 0.3" and 0.4" and
a few with FWHM < 0.3".
4.1.2. Dome Ventilation System
The main Dome Ventilation Sytem (DVS) was in general use by the spring of
1998, when 13 of its 16 apertures were fully functional. The remaining
three require heavy mechanical (crane) work to correct defects, and at
the time of writing are not yet operational.
Nevertheless, as indicated in last year's report, the average
inside-outside temperature differences are now close to zero for a
significant fraction of the time (see below). The DVS has proved
reliable in general use but is vulnerable to jamming if ice forms on the
guides of the roller doors, and quite careful procedures must be followed
to avoid this.
In mid-1998, after a review of the dome thermal behaviour by a Co-op
student, Patti Smith, fans were installed on the South column platform.
These were pointed upwards so as to circulate air through the top of the
dome, to prevent heated air accumulating there and heating the top-end
steelwork. The plant room ventilation system was set to extract air (at
around 10 volumes/hour) from the dome, which it does at floor level,
thereby removing any warmed air moving down from the upper parts of the
dome. The measured circulation rates and the lack of any temperature rise
in the lower structures indicated that this strategy worked well.
For the experiment, during the month of September, the fans were switched
on every second day and the averaged results from ``fans on'' and ``fans
off'' days compared. The fans reduced the warming of the top-end by about
a third and reduced the average inside-outside temperature difference
from about half a degree to approximately zero for more than half the
night. On the nights with the fans on the image FWHM averaged 0.04"
smaller than when the fans were off, a 15% improvement in resolution.
Further experiments with increased fan power have been conducted in 1999.
4.1.3. Dome Floor Insulation
As outlined in the previous report, safety considerations led to the
abandoning of plans to insulate the upper surface of the concrete dome
floor with removable foam-plywood panels. Instead, the underside of the
floor was insulated to minimise tranfer into the concrete, and hence
the dome air, of heat from the crew rooms, etc., below. At the same time
changes were made to the downstairs ceilings and doors to improve the
confinement of heat to the areas in which it is needed.
The design and management of the project were largely carried through by
Patti Smith. The installation was completed early in 1999.
4.1.4. Primary Mirror Cooling
Further progress was made with the installation of the cooling system as
effort allowed. By the end of 1998 detailed design was complete and
installation of coolant plumbing was well advanced. This continued into
1999.
4.1.5. Secondary Mirror Replacement
As noted last year, the new secondary mirror displayed several defects,
principally a turned-down edge, which reduced the limiting achievable
Strehl ratio to around 70%; print-through of the light-weighting pattern,
which makes the use of Adaptive Optics difficult or impossible; and
trefoil aberration induced by thermal stress at the mounting points,
further reducing the limiting Strehl ratio and, with the
light-weighting pattern, generating a hexagonal pedestal under the images.
In 1997 an agreement was accordingly reached with the MPIA for the
provision of a new secondary mirror without these defects.
In early 1998 a careful review of potential suppliers was undertaken to
establish technology availability, risks and likely costs. This
established that obtaining a new mirror from a single supplier in the USA
would certainly be expensive, probably between $150 k and $200 k, i.e. at
least twice what the MPIA estimated their managed procurement would cost,
and therefore involving significant costs (50% or more of the total) for
PPARC.
After careful consideration of risks and costs the MPIA were authorised
to proceed with the procurement as proposed. Fabrication of the slightly
oversized optic, testing through the rear surface, followed by reduction
to the specified diameter to remove residual turned-down edge (TDE) on
the optical surface, was carried out by Horst Kaufmann Prazisionsoptik,
Crailsheim-Wittau, Germany, in late 1998. In early 1999 the mirror was
lightweighted by grinding at BNM GmbH, Jena, Germany, and the back
surface was then acid-etched for stress relief by Carl Zeiss, Oberkochen,
to eliminate print-through of the lighweighting pattern, which is
prominent in the original mirror. Finally, new athermal mounts were glued
into the appropriate pockets in the back of the mirror. These were
designed by Physik Instrumente, Waldbronn, Germany, and R.J.Bennett at
the UKATC, to eliminate the thermal stresses which at normal operating
temperature impose significant R5 and R3 trefoil aberrations on
the original secondary. 1
|