|
UKIRT Annual Report 2000
THE UNITED KINGDOM INFRARED TELESCOPE
ANNUAL REPORT
2000
3.4. Instruments
3.4.1. CGS4 (1-5 micron multiple-resolution spectrometer with
256 256 array)
CGS4 was used for 44% of the PATT nights in semesters 00A and 00B,
and 2000 was another successful year for this instrument, with only
0.3% of clear time lost to faults.
In March, a new A/D converter was installed which fixed a minor but
long-lived CGS4 problem: one channel was consistently dropping bits. Read
noise was consistently excellent (in the mid-to-high 30-electron range)
after this change was made. In July the echelle grating was
replaced with the 150 l/mm grating, internal parts of the coldhead were
replaced, the new CVFs were inspected and the grating mechanism was also
inspected for brake wear. No problems were found and the engineering was
completed smoothly.
CGS4 operated well under the new ORAC system with only a few minor
problems, mostly involving aperture definitions and peakup. Initial ORACDR
recipes for data reduction covered the basic observing modes of CGS4, and
these proved to be ample for most purposes. The only notable data
reduction issue was the discovery that for some months, the pipeline
flux calibration routine had been incorrectly implemented. This was fixed
shortly after it was discovered, the major
surprise being the long delay before the problem was noticed. All affected
observers were advised to re-calibrate their data off-line.
3.4.2. IRCAM/TUFTI (1-5 micron camera with 256 256 array)
IRCAM was used for 12% of the PATT nights in semesters 00A and 00B.
Half of the runs used IRPOL. All runs in semester 00A were affected by the
persistent data-transfer problem known as ``BDS errors''. On the last
IRCAM/TUFTI run of the semester, a hard failure of
the waveform generator board resulted in continuous BDS errors. The board
was replaced with the spare which after some work
appeared to have finally cured the problem. However, when
winter returned the errors also returned. In late January 2001 the
cause was finally traced to the cooling fans in the ALICE rack, and
investigations were beginning into how the fan arrangement might be
altered.
Mechanically, IRCAM performed well through the year, with only two
cryogenic incidents. In October/November the cryostat was warmed up and
maintenance work was carried out on the filter wheels, which had been
producing datum errors. The work was successful, reducing the lost time
from more than two hours per semester to 0.6.
Transfer of IRCAM/TUFTI data acquisition over to the ORAC system was
initiated in the first half of 2000, and successfully introduced to
outside users at the start of 00B. Software changes were made to the new
TCS and the new ORAC data acquisition system that
made the instrument more robust and convenient to use.
3.4.3. UFTI (1-2.5 micron camera with 1024 1024 array)
UFTI was used for 38% of the PATT nights in semesters 00A and 00B and
was the first instrument to achieve robust performance under
ORAC. Initial problems with data acquisition and handling were
quickly resolved.
Documentation on the use of UFTI with ORAC was quickly made
available on the UKIRT web pages. Links to the recipe descriptions are in
fact links to cgi scripts which automatically generate the online
documentation from the perl source comments; thus the
documentation will never become out of step with the actual running systems.
A programme was begun to monitor bad pixels and to regularly update the
UFTI bad-pixel mask used by ORAC-DR; in the initial state, 0.36% of the 1
million pixels in the UFTI array were flagged as ``bad'' by the bad-pixel
mask. Initial results of the study indicated that, over a period of a few
weeks, some pixels ``migrate'' across the good/bad cut-off; but that the
bad-pixel mask nevertheless remains accurate to within 5%.
High humidity continued to cause condensation on the cryostat
window, and a significant fraction of the small amount of UFTI time lost
was to this cause. A fan was installed in front of the UFTI window
in mid-December to aid in clearing condensation, and worked well
immediately. The fan is always on; a heater is also available to remove
condensation in need. Its use is not an option with
the Fabry-Perot, which is highly temperature-sensitive.
Towards the end of 00B, low level noise was seen and the array controller
was swapped for the spare early in 2001. Some vignetting of the filters
was seen in 00B and in March 2001 the instrument filter wheels were
refurbished, solving the problem. After a power outage in November UFTI
had cooling problems which were traced to a faulty coldhead, which was
replaced.
Engineering was scheduled for early October when a number of filters would
have been upgraded to ``Mauna Kea'' standards. The H 2-1S(1) filter, which
gave elongated images, should also have been replaced. In the event,
Infrared Engineering - the UK-based company contracted by JAC (and a
number of other observatories) - were unable to deliver these filters and
the engineering was postponed.
3.4.4. Accessories
UKIRT's support for polarimetry options on all instruments is unique
among MK facilities. IRPOL, the facility polarimetry unit, also found
favour with the visitor instrument TRISPEC. During the course of 2000,
IRPOL was integrated with the new telescope control and data acquisition
systems, and the first observer to use ORAC did so with IRCAM3
and IRPOL at the very beginning of Semester 00B. A number of ORAC template
sequences and data reduction recipes are now offered for IRPOL imaging.
The (K-band) Fabry-Perot was used with UFTI for three projects in 00A and
a further three in 00B. A simple shell script is now being used to speed
up the alignment process. In addition, with the DVS,
stable temperature conditions are reached within an hour or so of the
start of each night. FP control with ORAC was commissioned in October;
data reduction recipes and associated ORAC template observing sequences
are now available. Use of UFTI+FP with
ORAC is fully documented on the UKIRT web pages.
Actively cooling and/or maintaining the 400 km/s FP (and indeed the
higher-resolution FPs) at a fixed temperature might be an option if
the FP and UFTI are to be offered after the arrival of UIST.
3.4.5. Visitor Instruments
Visitor instruments MAX and TRISPEC were used for 6% of the PATT nights
in semesters 00A and 00B. Both performed quite well.
TRISPEC was allocated time at the end of August, just post-ORAC
commissioning. Two PATT runs independent of the TRISPEC team were carried
out, in addition to observations by Chrysostomou and the TRISPEC team and
an engineering night allocated at the start of the run. The observing went
reasonably well, and it was hoped that a further visit could be undertaken
once some flexure issues were cured.
MAX, the MPIA thermal-IR camera, had a productive visit in December 2000.
Images of Orion at 10 and 20 micron were completed once some instrument
problems had been identified and cured. The thermal stability of UKIRT was
praised by the MPIA team; this is a good sign for future use with
Michelle.
|