The focus excursions measured by the autofocus system are a measure of
the seeing: standard atmospheric theory predicts that the
seeing-induced
image diameter is linearly related to the RMS atmospheric-induced focus
fluctuations. In this mode the fast guider is acting as a form of
Differential Image Motion Monitor (DIMM), the standard site-appraisal
seeing monitoring systems. Zrms should therefore be a robust
and invaluable measure of the seeing (as opposed to the
delivered image quality!) at UKIRT.
We are in the process of exploring the possibility of using
Zrms to derive reliable routine measures of the delivered
image
quality, which is an important parameter for flexible scheduling
purposes,
and are currently attempting to calibrate the system. An approximate
calibration, which is reasonably accurate when the UFTI image FWHM
~0."6, but certainly pessimistic when it is <0."4, is
currently in use for rough image-size estimates, e.g. when the
spectrometer is in use.
|